Performing the CAL-V™ Calibration Validation Test
i
NOTE!
The FT4X will stop measuring flow when performing this test.
Press
FLO (F2)
from the Main Menu. The display will show:
F1 F2 F3 F4
Press
DGN (F1).
The display will show:
F1 F2 F3 F4
Press
CAL-V (F2).
The display will show:
F1 F2 F3 F4
Press
VER (F1)
to perform the CAL-V™ verification test.
F1 F2 F3 F4
FLOW MENU 1
DGN UNT FM2 EXIT
DIAGNOSTIC MENU
SIM CAL-V EXIT
CAL-V MENU
VER EXIT
VERIFY CAL-V?
YES NO
66
Model FT4X
OPERA
TION
Operation