ONLINE ELECTRONICS LTD
ID5000A_5001_A03
Page 10 of 21
3.3.16.
SET TRIG DELAY
Accessed from SETUP MODE 1. This option sets a time delay after detection. As the unit will
detect debris and gas being pushed in front of the PIG, the unit continues to monitor the pipe
for a PIG passage during the TRIG DELAY period. If a trigger is received during this time the
delay will be reset to ensure the recorded event time is more accurate.
3.3.17.
SET FILTER
Accessed from SETUP MODE 1. If there is a poor signal path through the pipeline due to gas or
solids in the pipeline product then the FILTER value may be increased to avoid the possibility
of false triggers. It may be adjusted to any value from 001 to 255. See section 3.6 DETECTION
ALGORITHM for more information.
3.3.18.
SET CONFIG
Accessed from SETUP MODE 1. The SET CONFIG menu allows the user to enable special
functions within the unit. The number the CONFIG option is chosen by adding up the values for
the desired options in the table below.
Option
Number
Effect when selected
Invert Relay Operation
1
Relay will be normally closed and will open when
signalling
Debug display
2
Displays information useful for debugging on the
OLED when something is detected in the pipe.
LED Flash disable
4
Disables the flashing LEDs while listening. The
LEDs will still flash when a PIG is detected
Rechargeable Battery
8
Enables unit to recognise rechargeable battery
fitted and to adjust battery monitor accordingly
For example if you want the relay option to be inverted and the LED flash disabled:
Number for inverting relay
= 1
Number for Debug display
= 0
Number for LED flash disable
= 4
CONFIG number = 1 + 0 + 4 = 5
There for the number to be input to the CONFIG menu is 5.