ONLINE ELECTRONICS LTD
ID5000A_5001_A03
Page 8 of 21
3.3.3.
DELETE ALL EVENTS
Accessed from DETECTION MODE 0. Selecting this item will display the message *WARNING*
PRESS TO ERASE ALL EVENTS. If the user wishes to delete all logged data then they should
press the switch once before the counter reaches zero. If they do not wish to erase all logged
data then simply allow the counter to reach zero without pressing the switch.
3.3.4.
MODE1 SETUP
Accessed from DETECTION MODE 0. Selecting this item leaves DETECT MODE 0 and enters
SETUP MODE 1. See section 3.4 SETUP MODE 1.
3.3.5.
SHUTDOWN
Accessed from DETECTION MODE 0. Selecting this item switches off the unit. When the
counter reaches zero the display will go blank and the LEDs will flash once. All logged events
and parameters will still be available when the unit is next switched on.
3.3.6.
SAVE GAIN AND PW
Accessed from SETUP MODE 1. This saves the current GAIN and PULSE WIDTH values
configured from within SETUP MODE 1. This should be done whenever any changes are made
to the GAIN and PULSE WIDTH values from within SETUP MODE 1 which are to be stored. If
this is not done then any changes to GAIN and PULSE WIDTH will be lost if SETUP MODE 1 is
exited.
3.3.7.
MODE0 DETECT
Accessed from SETUP MODE 1. Selecting this item leaves SETUP MODE 1 and enters DETECT
MODE 0. If the SAVE GAIN AND PW item was not selected prior to this then any changes to
their values will not be saved.
3.3.8.
ADJUST GAIN UP
Accessed from SETUP MODE 1. This item changes the function of the switch to increment
GAIN. The unit will return to SETUP MODE 1 after choosing this option and every short press of
the switch will increment the gain value. Once GAIN is set the SAVE GAIN AND PW option
should be selected.
3.3.9.
ADJUST GAIN DOWN
Accessed from SETUP MODE 1. This item changes the function of the switch to decrement
GAIN. The unit will return to SETUP MODE 1 after choosing this option and every short press of
the switch will decrement the gain value. Once GAIN is set the SAVE GAIN AND PW option
should be selected.
3.3.10.
AJST PULSE WIDTH
Accessed from SETUP MODE 1. This item changes the function of the switch to adjust PULSE
WIDTH. The unit will return to SETUP MODE 1 after choosing this option and every short press
of the switch will increment the pulse width value. Once the pulse width is set the SAVE GAIN
AND PW option should be selected.