HOBO U30 Station (NRC) User’s Guide
49
Testing Smart Sensors
Use HOBOware Pro if you need to verify the accuracy of the Smart Sensors,
which is recommended on a yearly basis, or if you need to isolate a Smart
Sensor that is generating an error.
1.
Connect the HOBO U30 Station to the computer using the USB
cable. If it is currently logging, stop logging through HOBOware
Pro.
2.
Unplug all Smart Sensors if you are troubleshooting an error. You
can leave them plugged in if you are not getting any sensor errors
and you are just checking them for accuracy.
3.
Plug one Smart Sensor back in. From the Launch dialog in
HOBOware Pro, check the current reading and make sure it is
accurate. Repeat with the remaining Smart Sensors.
50