background image

 

HOBO External Temp/RH Data Logger (UX100-023A) Manual 

1-800-LOGGERS 

www.onsetcomp.com

 

on the LCD screen. Note that the alarm limits are only 
checked when the logger’s LCD screen refreshes every 15 
seconds. 

 

The actual values for the high and low alarm limits are set 
to the closest value supported by the logger. For 
example, the closest value to 85°F that the UX100 series 
logger can record is 84.990°F and the closest value to 
32°F is 32.043°F. In addition, alarms can trip or clear 
when the sensor reading is within the logger 
specifications of 0.02°C resolution. This means the value 
that triggers the alarm may differ slightly than the value 
entered. For example, if the High Alarm is set to 
75.999°F, the alarm can trip when the sensor reading is 
75.994°F (which is within the 0.02°C resolution). 

 

When you read out the logger, high and low alarm levels 
will be displayed on the plot along with “Chan <#> Alarm 
Tripped” and “Chan <#> Alarm Cleared” events showing 
when the sensor alarm tripped and cleared. The “Chan 
<#> Alarm Cleared” event contains the value that was 
furthest out of range for the sensor before the alarm 
cleared (see the Points table for the actual value). 

Burst Logging 

Burst logging is a logging mode that allows you to set up more 
frequent logging when a specified condition is met. For 
example, let’s say the logger is recording data at a 5-minute 
logging interval and burst logging is configured to log every 10 
seconds when the temperature goes above 85°F (the high limit) 
or falls below 32°F (the low limit). This means the logger will 
record data every 5 minutes as long as the temperature 
remains between 85°F and 32°F. Once the temperature reaches 
90°F, for example, the logger will switch to the faster logging 
rate and record data every 10 seconds until the temperature 
falls back below the high limit (or 85°F in this case). At that 
time, logging then resumes every 5 minutes at the normal 
logging interval. Similarly, if the temperature falls to 30°F, for 
example, then the logger would switch to burst logging mode 
again and record data every 10 seconds. Once the temperature 
rises back to 32°F, the logger will then return to normal mode, 
logging every 5 minutes. 

To set up burst logging: 

1.

 

Select Burst for Logging Mode in the Launch Logger 
window. If Burst has already been configured for this 
logger, click the Edit button in the Launch Logger 
window. Note that you cannot set up alarms with burst 
logging. You also cannot select the Stop Logging option 
“Never (wrap when full)”” if burst logging is configured. 

2.

 

In the Burst Logging window, select a sensor. In the 
example at right, the temperature sensor was selected. 

3.

 

Select the High Limit checkbox if you want to set up a 
condition in which burst logging will occur when the 
sensor reading rises above the high limit value. Type in 
the value or drag the red upper slider (see example 
below). 

4.

 

Select the Low Limit checkbox if you want to set up a 
condition in which burst logging will occur when the 
sensor reading falls below the low limit value. Type in the 
value or drag the blue lower slider. 

 

5.

 

Repeat steps 2 through 4 for the other sensor, if desired. 

6.

 

Set the burst logging interval, which must be less than 
the logging interval. Select either a preset burst logging 
interval or select Custom and enter your own interval. 
Keep in mind that the more frequent the burst logging 
rate, the greater the impact on battery life and the 
shorter the logging duration. 

7.

 

Click OK when done. This will return you to the Launch 
Logger window. Click the Edit button next to Logging 
Mode in the Launch Logger window to make additional 
changes. 

Notes: 

 

Once the logger is launched, the high and low burst 
logging limits are only checked when the logger’s 
LCD screen refreshes once every 15 seconds. Therefore, 
if you set the logging interval to less than 15 seconds and 
the sensor reading falls outside the limits, the burst 
logging will not begin until the next 15-second refresh 
cycle. 

 

If high and/or low limits have been configured for more 
than one sensor, then burst logging will begin when any 
high or low condition goes out of range. Burst logging will 
not end until all conditions on all sensors are back within 
normal range. 

 

The actual values for the burst logging limits are set to 
the closest value supported by the logger. For example, 
the closest value to 85°F that the logger can record is 
84.990°F and the closest value to 32°F is 32.043°F. 

 

Burst logging mode can begin or end when the sensor 
reading is within the logger specifications of 0.02°C 
resolution. This means the value that triggers burst 
logging may differ slightly than the value entered. For 
example, if the high limit for a temperature alarm is set 
to 75.999°F, burst logging can start when the sensor 
reading is 75.994°F (which is within the 0.02°C 
resolution). 

 

Once the high or low condition clears, the logging 
interval time will be calculated using the last recorded 
data point in burst logging mode, not the last data point 
recorded in “normal mode.” For example, let’s assume 

Summary of Contents for Hobo UX100-023A

Page 1: ...ical Resolution 0 01 Response Time 15 seconds to 90 in airflow of 1 m s 2 2 mph Drift 1 per year typical Cable Length 1 83 m 6 ft Logger Logger Operating Range Logging 20 to 70 C 4 to 158 F 0 to 95 RH...

Page 2: ...oximate battery power remaining The logger has been configured to stop logging when memory fills The memory bar indicates the approximate space remaining in the logger to record data When first launch...

Page 3: ...p logging and choosing a logging mode 1 Connect the logger and open the Launch Logger window To connect the logger to a computer use the USB cable provided Click the Launch icon on the HOBOware toolba...

Page 4: ...g Out the Logger for details Setting up Alarms You can set an alarm to trip on the logger when a sensor reading rises above or falls below a specified value To set an alarm 1 Click the Alarms button f...

Page 5: ...th burst logging You also cannot select the Stop Logging option Never wrap when full if burst logging is configured 2 In the Burst Logging window select a sensor In the example at right the temperatur...

Page 6: ...Standard Deviation checkboxes for each of the statistics you want to calculate during logging Note that Average is automatically enabled when selecting Standard Deviation Important Statistics apply t...

Page 7: ...if burst or statistics logging modes are active and battery performance A new battery typically lasts 1 year with logging intervals greater than 1 minute Deployments in extremely cold or hot temperat...

Reviews: