background image

 

InTemp CX400 Series Temperature Logger Manual 

1-508-759-9500 (U.S. and International) 4 

www.onsetcomp.com 

1-800-LOGGERS (U.S. only) 

with the InTempVerify app, then you must create a logger 
profile with InTempVerify enabled. See 
www.intempconnect.com/help for details. 

InTemp app only users:

 The logger includes preset profiles. 

To set up a custom profile, tap the Settings icon and tap 
CX400 Logger. Also, if you need to perform daily logger 
checks, tap Record CX400 Logger Checks under Settings and 
select Once Daily or Twice Daily. This should be done before 
configuring the logger. 

See www.intempconnect.com/help for details on setting up 
custom profiles in both the app and InTempConnect and on 
setting up trip information. 

a.

 

Tap the Devices icon in the app. Find the logger in the list 
and tap it to connect to it. 

If you are having trouble connecting: 

 

Make sure the logger is within range of your mobile 
device. The range for successful wireless 
communication is approximately 30.5 m (100 ft) with 
full line-of-sight. 

 

If your device can connect to the logger intermittently 
or loses its connection, move closer to the logger, 
within sight if possible. 

 

Change the orientation of your phone or tablet to 
ensure the antenna in your device is pointed toward 
the logger. Obstacles between the antenna in the 
device and the logger may result in intermittent 
connections. 

 

If the logger appears in the list, but you cannot connect 
to it, close the app, power down the mobile device, 
and then turn it back on. This forces the previous 
Bluetooth connection to close. 

b.

 

Once connected, tap Configure. Swipe left and right to 
select a logger profile. Type a name or label for the 
logger. Tap Start to load the selected profile to the 
logger. 

InTempConnect users:

 If trip information fields 

were set up, you will be prompted to enter additional 
information. Tap Start in the upper right corner when 
done. 

5.

 

Deploy and start the logger

. Deploy the logger to the 

location where you will be monitoring the temperature. 
Logging will begin based on the settings in the profile 
selected. If the logger was configured to perform daily 
checks, connect to the logger and tap Perform (Morning, 
Afternoon, or Daily) Check every day.

 

Once logging begins, the logger will display the current 
temperature reading and the minimum and maximum readings 
within the current 24-hour period (midnight from one day to 
midnight the next day) if the logger was set to record logger 
checks (see 

Performing Logger Checks

). Otherwise, the 

minimum and maximum readings represent the entire logging 
period and only reset when the logger is downloaded and 
restarted or stopped and reconfigured (they will not reset if you 
download the logger and continue logging). These minimum 
and maximum values are also available in the logger reports 
(see 

Downloading the Loggers

). 

For CX402 models, you can click Next to switch between external 
probe and ambient temperature readings. Minimum and 
maximum readings are only available for the external probe. 

Logger Alarms 

There are three conditions that can trip an alarm: 

 

When the temperature reading by the external probe (if 
applicable) or the ambient temperature is outside the 
range specified in the logger profile. 

 

When an external probe (if applicable) is disconnected 
during logging. 

 

When the logger battery drops to 15% or the equivalent 
of one bar on the LCD battery icon. 

You can enable or disable alarms and set temperature alarm 
thresholds in logger profiles that you create in either 
InTempConnect or in the app. 

When a temperature alarm trips: 

 

The logger LED will blink every 5 seconds. 

 

The alarm icon will appear on the LCD and in the app. 

 

The logger will beep once every 15 seconds (unless 
audible alarms are disabled in the logger profile). 

 

An Alarm Tripped event is logged. 

When an external probe is removed: 

 

The logger LED will blink every 5 seconds. 

 

“ERROR” and “PROBE” will appear on the LCD and 
“ERROR” will appear in the app. 

 

An alarm icon will appear in the app. 

 

The logger will beep once every 15 seconds. 

 

A Probe Disconnected event is logged. 

When a low battery alarm trips: 

 

The battery icon on the LCD will flash. 

 

The logger will quickly beep three times every 15 
seconds. 

 

A Low Battery event is logged. 

To mute a beeping alarm, press the Mute button on the logger. 
Once muted, you cannot turn the beeping back on. Note that if 
a temperature and/or probe alarm is occurring at the same 
time as a low battery alarm, pressing the mute button will 
silence all alarms. 

Download the logger to view details about a tripped alarm and 
to clear temperature alarm indicators in the app and on the 
LCD (the probe must be reconnected for ERROR to clear on the 
LCD). For a temperature alarm, the tripped alarm will clear 
once the logger is downloaded and restarted. Replace the 
batteries in the logger to clear a battery alarm. 

Note:

 Download 

the logger before replacing the batteries to ensure no data is 
lost. 

Passkey Protection 

The logger is protected by an encrypted passkey automatically 
generated by the InTemp app for InTempConnect users and 
optionally available if you are using the InTemp app only. The 
passkey uses a proprietary encryption algorithm that changes 
with every connection. 

Summary of Contents for InTemp CX402-T205

Page 1: ...o 212 F Resolution 0 024 C at 25 C 0 04 F at 77 F Drift 0 1 C 0 18 F per year NIST Calibration CX40x Txx and CX402 BxM Single point NIST calibration probe and logger body CX40x VFCxxx Single point NIS...

Page 2: ...or measures ambient temperature External Temperature Probe This is the probe that is plugged into the logger for measuring temperature CX402 models only Audible Alarm Speaker This is the speaker for t...

Page 3: ...oftware where you can monitor CX400 series logger configurations and view downloaded data online Using the InTemp app you can configure the logger with your phone or tablet and then download reports w...

Page 4: ...nimum and maximum readings represent the entire logging period and only reset when the logger is downloaded and restarted or stopped and reconfigured they will not reset if you download the logger and...

Page 5: ...empConnect when you download the logger Log in to InTempConnect to build custom reports requires privileges In addition InTempConnect users can also download CX loggers automatically on a regular basi...

Page 6: ...h this reset Logger Events The logger records the following events to track logger operation and status These events are listed in reports downloaded from the logger Event Name Definition Started The...

Page 7: ...ts in extremely cold or hot temperatures or a logging interval faster than 1 minute can impact battery life Estimates are not guaranteed due to uncertainties in initial battery conditions and operatin...

Page 8: ...o radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception wh...

Reviews: