P a n e l P C
M A T E - 1 9 0 1 / 2 2 0 1 / 2 4 0 1
16. If any of the following situations arises, get the equipment
checked by service personnel:
a. The power cord or plug is damaged.
b. Liquid has penetrated into the equipment.
c. The equipment has been exposed to moisture.
d. The equipment does not work well, or you cannot get it
to work according to the users manual.
e. The equipment has been dropped and damaged.
f. The equipment has obvious signs of breakage.
17. DO NOT LEAVE THIS EQUIPMENT IN AN UNCONTROLLED
ENVIRONMENT WHERE THE STORAGE TEMPERATURE IS
BELOW -20° C (-4°F) OR ABOVE 60° C (140° F). IT MAY
DAMAGE THE EQUIPMENT
.
18. External equipment intended for connection to signal
input/output or other connectors, shall comply with
relevant UL / IEC standard (e.g. UL 1950 for IT equipment
and ANSI/AAMI ES 60601-1: 2005 AND CAN/CSA-C22.2 No.
60601-1:08 / IEC 60601 series for systems – shall comply
with the standard IEC 60601-1-1, Safety requirements for
medical electrical systems. Equipment not complying
with UL 60601-1 shall be kept outside the patient
environment, as defined in the standard.
Caution:
It may cause the danger of explosion if battery is incorrectly
replaced. Replace only with same or equivalent type
recommended by the manufacturer.
Summary of Contents for MATE-1901
Page 15: ...Panel PC M A T E 1 9 0 1 2 2 0 1 2 4 0 1 General Information Chapter 1...
Page 24: ...Panel PC M A T E 1 9 0 1 2 2 0 1 2 4 0 1 Chapter 1 General Information 1 9 MATE 2201...
Page 25: ...Panel PC M A T E 1 9 0 1 2 2 0 1 2 4 0 1 Chapter 1 General Information 1 10 MATE 1901...
Page 34: ...Panel PC M A T E 1 9 0 1 2 2 0 1 2 4 0 1 Appendix A Miscellanea A 1 Miscellanea Appendix A...