Servicing and Maintenance
Servicing work must only be carried out by persons who have at least read the
safety regulations and are qualified according to the MPBetreibV (Operators of
Medical Products Ordinance) § 4 and 6.
A technical check and/or safety inspection must be conducted at least every
6 months and after a lengthy brake in use and before each further use. Refer
to section 7 for further detail. Any defects, damage or signs of wear must be
rectified without delay. Only original spare parts from Opera may be used,
otherwise all guarantees or warranties will be excluded.
Please check all fixings on your bed at least once a month. Regular cleaning and
maintenance will reveal any loose or worn parts.
5.6
Accessories
Use only mattresses compatible with the supplied side rails and bed. The
dimension between the mattress upper surface in an uncompressed condition
and the top edge of the upper side rail must be 220mm minimum. If this
dimension is less than 220mm, an extension side rail kit should be fitted.
5.7
Electromagnetic Compatibility
Regarding the emitted interference and interference resistance, the electric
motor units comply with the requirements of EN 60601-1-2:2007. However it is
possible that electrical devices can interfere with each other. In this case, switch
off the care bed for a short time or remove the interference source.
5.8
Storage and Handling
The bed should be stored between -10°C and 60°C, in a room with a relative
humidity between 30% and 75%, keep the bed dry and out of direct sunlight.
Attention: In the event of disinfection by large scale spraying with products
containing alcohol, there is a danger of explosion and fire.