31
|
Opera Care
Solo Bed Installation Guide and Technical Specification | Version 1
Item
Instruction for testing
Comment
Yes No
17.
Height of side rails above the mattress
at least 220mm?
18.
Height of side guards above the
mattress at least 220 mm?
19.
Have castors including locking brake
been tested for safe functioningg?
20.
Mains cable, connecting cables and
plugs without damage?
21.
Fixture available for safe transportation
of mains plug?
22.
Strain relief of the mains cable and
handset securely attached?
23.
Are all plug-in connections securely
attached? (Washers without damage?)
24.
Are cables laid correctly and safely?
(No damage)
25.
Motor housing and SMPS housing,
mains plug housing without damage?
26.
Are the thrust pipes of the height
adjustment motors undamaged?
27.
Functional test of the handset: can the
buttons be operated properly?
28.
Functional test of handset locking
device: On/Off working correctly?
29.
Testing of initial fault safety by means
of integrated blocking box in handset
30.
9V block battery OK / expiry date
sufficient until next test?
31.
Is the safe working load adhered to?
Comments: ..............................................................................................................................................
Place / Date: ............................................................................................................................................
Inspected by: ...........................................................................................................................................
Next inspection:......................................................................................................................................
Signature: .................................................................................................................................................
7.
Servicing