PERFORMING NON-UNIFORMITY CORRECTION (NUC)
The NUC process corrects for image noise or artifacts caused by external or
internal temperature changes.
To perform the NUC process:
1. Mount the lens cover (A) on the camera lens.
2. Press the
NUC button
(B) for 3 seconds.
The image on the screen freezes for a few seconds, indicating NUC is
in progress.
Perform the NUC process separately on either of the following:
z
IR image –
run NUC in normal or enhanced mode.
z
Thermography image –
run NUC in thermography mode.
NOTES
▪
Only perform NUC when there is image degradation.
▪
The NUC process does not resolve poor image quality caused by lens
condensation or dirt on the lens.
A
B
press for
3 seconds
Operation | Frequently Used Operations
41.
MENU
EyeCGas® 2.0 | User Manual Rev 01 | Doc P/N: UM-8G9A0000B