Calibration, Maintenance, and Troubleshooting
StarBright User Manual
100
tested periodically over their entire range. The test is performed with a high power laser that can
cover the entire detector range using a beam splitter and lower power sensor that has previously
been tested for linearity for comparison. Thus, in all cases, the linearity of the detectors is
traceable to NIST measurements.
Figure
14-2 Absorption vs. Wavelength of Various Pyroelectric Sensor Absorbers
Method of Calibration
The absorption of the various Ophir Thermopile absorbers can vary from disc to disc. Therefore, all
Ophir absorbers are individually calibrated against NIST traceable standards.
Total Accuracy of Calibration
A detailed discussion of Ophir calibration accuracy is available on the Ophir website at
14.2.4
Updating Thermopile Sensor Calibration Factors
To update Thermopile sensor calibration factors:
1.
Press
Menu
and select
Calibration
. The screen displays the following general message:
Figure
14-3 General Calibration Information Screen