Calibration, Maintenance, and Troubleshooting
StarBright User Manual
107
14.3.3
Updating Photodiode Sensor Calibration Factors
Photodiode sensors only have an overall power calibration factor. Changes to this factor at one
wavelength will affect readings at all other wavelengths proportionately.
Note:
Sensors equipped with a removable filter have two factors, one for each setting. Changing
the factor for one setting has no effect on power readings in the other settings.
To update Photodiode sensor calibration factors:
1.
Press
Menu
and select
Calibration
. The screen displays the following general message:
Figure
14-22 General Calibration Information Screen
2.
Press
Continue
. The screen displays the following Photodiode sensor-specific message:
Figure
14-23 Information Screen for Photodiode Sensors
3.
Press
Continue
to activate calibration. The main Measurement screen appears, displaying the
Calibration Factor value (on the left).
Figure
14-24 Calibration Activated on Main Measurement Screen