Calibration, Maintenance, and Troubleshooting
StarBright User Manual
111
2.
Press
Continue
. The screen displays the following Pyroelectric sensor-specific message:
Figure
14-29 Information Screen for Pyroelectric Sensors
3.
Press
Continue
to activate calibration. The main Measurement screen appears, displaying the
relevant
Overall
(or
Laser
)
Factor
value (on the left).
Overall Factor
: Available for all Pyroelectric sensors. Changes the overall calibration factor
of the sensor. This will affect readings for all laser settings. There is an overall calibration
factor for each pulse width option available in the sensor. For sensors with a removable
diffuser, there are two sets of factors; one set for each diffuser setting.
Laser Factor:
Available for the BB series of sensors that have discrete wavelength
calibration. Modify readings for this laser only.
Figure
14-30 Calibration Activated on Main Measurement Screen
4.
Press
Modify
. The screen displays the
Original
(current) reading and factor (on the left side of
the screen), and
Updated
factor and its effect on the reading (on the right). The
Updated
box
is highlighted, indicating it is modifiable.
Figure
14-31 Modifying Calibration for Pyroelectric Sensors