Graphical Displays
StarBright User Manual
64
To set pass/fail tracking:
1.
In
Power
or
Energy
mode, press
Menu
and enter
Display
.
2.
Go to
Graph Type
and select
Pass/Fail
. Return to the main display screen.
3.
Press the
Upper
and
Lower
buttons to set tolerance limits. If reading is out of range, an
appropriate warning will be displayed on the screen.
9.6.1
Measuring Modes
The
Measuring
modes that Pass/Fail displays can be used with include:
Power
Energy
9.6.2
Upper/Lower Limits
You can set the Upper and Lower Limits for passing measurements. The initial default Upper Limit
is the maximum of the present range, and the initial default Lower Limit is 0 (until the first time
both values are set to something else).
To set the Upper and Lower Limits:
1.
Press the
Upper
or
Lower
key. The respective
Modify Upper/Lower
Limit
window opens (see
).
2.
User the right and left Navigation keys to select the digit to update, including the exponent
(E+
00
).
3.
Use the up/down Navigation keys to set the select digit to the desired value. (The limits can be
set from -9999 to 9999 E -15 to E +12.)
If the measurement is out of range, “OVER” or “UNDER” will be displayed on the screen. If the
Color Scheme has been set to one of the full color options, then this will be displayed in RED,
along with the reading itself (as shown in
Figure
9-25 Pyroelectric Pass/Fail - Modify Upper Limit