MAINTENANCE SCHEDULE
SOTR MASK MAINTENANCE SCHEDULE
SOTR FILTER MAINTENANCE SCHEDULE
9
1
Clean the mask with 70% isopropyl alcohol and let dry.
2
Check that the snaps are in working condition.
3
Clean the Velcro of dust and dirt by using pressurized air.
4
Visually check the snaps, exhalation valves, inhalation valve, and filter gasket for signs of damage.
5
Visually check the SOTR facepiece for any signs of damage.
6
Check that both harnesses work properly and do not have any damage.
7
Visually check the microphone and conduct operational checks of the microphone.
NOTE:
If any defects are found, the damaged components should be replaced before use.
1
Always start your training or operation with a new filter.
2
Once the filter becomes difficult to breathe through, discard the filter* and replace with a new one.
3
Visually check the filter gasket on the front of the yoke for any signs of damage.
4
Replace the filter gasket after 50 filter changes (or earlier if damage is observed).
NOTE:
*Dispose of the filter according to your unit’s recommendation. The filter will contain
hazardous elements and must be disposed of accordingly. Local, state, and/or federal disposal
regulations may apply.
Summary of Contents for SOTR
Page 18: ...18 OPERATOR NOTES...