9 Operation
11
9.2.1 Start Measurement
From the standby menu, a measurement is
started with the key
.
First, the spectrum is displayed.
Use ‘”Auto” mode to get the best integration
time or, in fixed mode use
and
keys to
adjust integration time.
Notes:
The optimal integration time, gain (L/H) and the saturation of the measurement range are dis-
played at the upper display edge. The integration time is displayed in ms. Example:
T 100: The integration time is 100 ms
Hint:
For optimum measurement results, the measuring range should be fully con-
trolled. Very good measuring results are achieved when the measuring range
is at least 50%.
The modulation is linear to the set integration time. Example
:
55/5000
T10 M20%: Measuring range is 20% used => Bad
T40 M80%: Measuring range is 80% used => Very good
Switch between spectral and radiometric view
by pressing
.
If storing measurements is selected via menu
„recording time“, the measurement is auto-
matically recorded if the radiometric view is
shown.