9 Operation
18
9.3 Settings
By selecting the menu item „Settings“, you are being directed to the settings menu of the
UVpad E.
In this menu, the following functions are
available:
-
Settings of the sensitivity
-
Selection of the effective function
-
Language settings
-
Settings of date and time
-
Reset to factory settings
9.3.1 Sensitivity
The UVpad E adjusts the sensitivity automatically, in order to ensure the optimal signal-noise
ratio. The sensitivity is hereby set via the sampling rate. With high irradiances, the sampling
rate is increased in order to not exceed the measurement range. With low irradiances the sam-
pling rate is decreased. As a result in belt conveyor systems erroneous measurements may be
recorded. Please follow the instructions on the display and / or set a fixed integration time as
sensitivity. Five pre-defined sensitivities are available.
In the setting „
automatically
“, the UVpad E
adjusts the sensitivity, once a minimum irra-
diance is existent. This setting is suitable for
most applications.
In the setting „
low
“, the sampling rate is al-
ways between 10 ms and 100ms. The UVpad
E can measure UV-systems with a very high
irradiance or significant irradiance modifica-
tions (e.g. with shutter operation or with very
high tape speed).
In the setting „
high
“, the UVpad E measures
with a sampling rate of 1000 ms. This is ideal
for applications with a low irradiance.
In the settings „fixed“ and „fixed (Res 0.01)“,
the integration time is predetermined. The
UVpad E can be used optimally for known
systems with shutters
Note:
In irradiation profile resolution is enhanced
from 1 mW/cm² to 0.01 mW/ m² for the set-
tings "high" and "Fixed (Res. 0.01). The
measuring range then is 0 .. 650 mW/cm².