12 Notes for optimal measurement results
32
creases. Therefore, the UVpad E has an automatic correction function. However, low UVC
values can possibly not be avoided, even if the UV lamp does not emit UVC.
12.4 The UVpad E displays “ - “
Just like each measurement device, the UVpad E has a fixed measurement range. If this range
is fallen below, then the UV radiation cannot be captured correctly. In that case, the UVpad E
displays “ – “.
Select a higher, lower or automatic sensitivity or set a long integration time.
12.5 The measured value is not reproducible
Measurements are only stored and displayed in UVpad E when the saturation is between 10%
and 100%. This can cause to saturation direct under the light source, however correct meas-
urement may be recorded before or after that. Especially UV LEDs can lead to a correct meas-
urement switching the LED on within one measurement, while all other are saturated.
Select a fixed integration time.
Note: Reduce the integration time to 10ms and not saturation. Then set the
highest possible integration time