13 Cleaning, Maintenance & Troubleshooting
33
13 Cleaning, Maintenance & Troubleshooting
13.1 Calibration
The calibration of the UVpad Es is checked during the initialization. If the UVpad E is not
calibrated properly anymore, then this is indicated by a note after the initialization process.
Measurements are also possible at any time with an expired calibration.
Confirm with the key
.
The UVpad E can only be maintained and calibrated by the manufacturer. For this purpose,
please consult your supplier.
UV spectroradiometer are in principle subject to a high high-energy load by
UV radiation. This can affect the calibration and lead to an early due date for
a re-calibration. This can depend on the individual use, the environment and
the radiation sources.
13.2 Cleaning
Skin oil and contamination are absorbing in the UV spectral range and thus affect the meas-
urement result. Finger prints or contamination on the optically active entrance opening are
therefore generally to be avoided.
Should a cleaning become necessary, please use a lintless cloth and Isopropanol. Clean the
optically active surface with circular movements. Avoid a strong force. If necessary, leave the
UVpad E to dry completely after cleaning it.
Please note:
For cleaning the optically active surface, please wear gloves.
Do not use Acetone or other chemical cleaning agents to clean the UVpad E.