If using higher power objectives, especially more than x10, you will notice the image to
be ‘tinny’ (lacking in good plain area contrast but having excessive edge-type contrast)
and / or dark. This is because you require diffuse light to enter at a larger range of
angles for higher power objectives and the best way to achieve this is with a condenser
module such as the optional PUMA Abbe condenser upgrade.
Another advantage of using the PUMA Abbe condenser upgrade module is that is comes
with a filter slot which allows you to shape the Fourier aperture of the condenser so you
can achieve advanced imaging modalities such as dark ground (dark field) microscopy
and Schlieren phase contrast. These can be achieved with simple fixed filters and do
not
require the use of the electronic spatial light modulator (which is a more advanced
optional extra). The next chapter gives more detail about optional upgrade modules.
PUMA Foundation Scope
Page 23 of 40
User’s Manual v1.0