Relay Setting
18
8
Relay Setting
Set the relay output setting from SC-U1.
The factory default setting is OFF for relay output.
To set the relay setting, press
.
Set the alarm to be output when the measurement value exceeds a certain
value.
The following settings are available.
Alarm output setting
Setting item
Setting range
Turbidity (NTU, FNU,
mg/L, no unit) [Turb.]
Turbidity (
度
)
[Turb.]
Trigger
High Alarm/Low Alarm
Active Level
0.00 to 1000.0
0.00 to 100.0
Hysteresis High
Alarm
0.00 to Active level
0.00 to Active level
Low
Alarm
0.00 to (1000.0 - Active
level)
0.00 to (100.0 - Active
level)
ON Delay
OFF
1 to 120 minutes
OFF Delay
OFF
1 to 120 minutes
The active level setting unit depends on the display resolution. (For display reso-
lution, refer to “15 Specifications” P.38)
For operation, refer to the "Universal Transmitter SC-U1 Operation
Manual", "5.4 Relay Setting" and "
◆
Alarm output setting".
Caution
Pressing
automatically retains the measurement value.
Caution
When “Display item (unit selection)” P.14 is changed, the
output item of the channel of relay setting that has set the display
item becomes OFF. Set again if necessary.