Adjustment
23
Adjust and display the measurement values according to usage. In addition, adjust
and display the measurement values according to the environment where used.
To adjust, press
.
Caution
・
Pressing
automatically retains the measurement value.
・
If offset, span, or 2-point adjustment has already been
performed, be sure to initialize the adjustment before performing
the 2-point adjustment.
・
When “Display item (unit selection)” P.14 is changed, the
values of the offset adjustment, span adjustment, and 2-point
adjustment that have the display item are reset to the factory
default values.
Set again if necessary.
10
Adjustment
Set the offset coefficient to add to a
measurement value.
Display the measurement value with
offset-adjusted to any display value.
For 4-20 mA output (“7 4-20 mA Setting” P.16)
and alarm output (“Alarm output setting” P.18),
the adjusted values are applied.
Offset adjustment
Measurement item (unit)
[Display]
Setting range
Factory default setting
Turbidity (NTU, FNU, mg/L,
no unit)
[Turb.]
-100.0 to 100.0 0.00
Turbidity (
度
)
[Turb.]
-10.0 to 10.0
0.00
Display value
after offset
adjustment
Measurement
value
Press
.
The measurement item selection display is displayed.
1