Adjustment
27
Measurement item
(unit)
[Display]
Measurement item setting range
(display value/adjusted value)
Factory de-
fault setting
Turbidity (NTU,
FNU, mg/L, no unit)
[Turb.]
[Point 1]: 0.00 to (2nd point value - 2)
0.00
[Point 2]: (1st point value + 2) to 1000.0 1000.0
Turbidity (
度
)
[Turb.]
[Point 1]: 0.00 to (2nd point value - 2)
0.00
[Point 2]: (1st point value + 2) to 100.0
100.0
Caution
If offset, span, or 2-point adjustment has already been performed,
be sure to initialize the adjustment before performing the 2-point
adjustment.
Select [2-Point Adjustment]
using
, and press
.
3
$GMXVW
㸰
7XUE76&'
2IIVHW$GMXVWPHQW178
6SDQ$GMXVWPHQW
3RLQW$GMXVWPHQW
,QLWLDOL]H$GMXVWPHQW9DOXH
Press
.
The measurement item selection display is displayed.
1
2
$GMXVW
0HDVXUHPHQW,WHP6HOHFW
㸯
S+ࠉ6(S+'
㸰
7XUE76&'
㸱
'2'26
㸱
7HPS'26
Measurement item
Sensor model
Sensor channel
Select [Turb. TSC-10-D] using
, and press
.
Use
to switch pages.
The adjustment menu is displayed.