35
Error code and details
Cause
Action
E2000
Sensor processing/sensor
The sensor cable
is not properly
wired.
Turn off power of the
transmitter, check the
wiring, and be sure to
wire properly.
The sensor is out
of order.
The sensor must be
repaired. Contact the
dealer.
E3100 to E3199
Measurement error
5/<
(
㸯
7XUE
The detection
window is not
clean.
Clean the detection
window if it is not clean.
E3200 to E3299
Temperature error
㸯
(
7XUE
5/<
178
The temperature
of the water to
measure is out of
the usage range.
The water temperature
must be 0 to 40
°
C.
E3300 to E3399
Humidity error
㸯
(
7XUE
5/<
178
The internal
humidity of the
unit exceeds its
limit.
The unit must be
overhauled or repaired.
Contact the dealer.
E4100 to E4199
Calibration error
&DOLEUDWLRQ
㸯
7XUE76&'
ࠉࠉࠉࠉ=(52&DOLEUDWLRQ
E4100
The detection
window is not
clean.
Clean the detection
window if it is not clean.