9.1 Light intensity adjustment
1.
Rotate the light adjustment ring
①
on the right side of the
microscope body. (Fig. 7)
As an alternative:
1.
Move the mouse button on the upper left corner of the screen.
2.
The Main Control Panel appears (see chapter 10.1 a).
3.
Adjust the illumination using the Camera Control Panel of the
built-in software.
9.2 Coarse focus tension adjustment
1.
To adjust the focusing tension, hold the left knob
②
and ro
-
tate the right knob
③
. (Fig. 8)
• Clockwise rotation increases the tension while counter-
clockwise rotation loosens it.
•
Adjust the focusing tension to appropriate position, to pre
-
vent the microscope body dropping during observation, and
also make focusing stable.
9.3 Parfocality adjustment
1.
Rotate the zoom adjustment ring
④
to the minimum magni
-
fication. (Fig. 9)
2.
While observing on the monitor, focus the specimen.
3.
Rotate the zoom adjustment ring
④
to the maximum mag
-
nification.
4.
Observe the image on the monitor: if it is not clear, rotate the
focusing knob to make it sharp.
5.
Return zoom adjustment ring to the minimum magnification.
Now parfocality adjustment should be set.
9.4 Focusing
Put the sample to be observed on the stand plate and focus the
sample using the focusing knobs.
•
If necessary adjust the height of the microscope head along
the pillar: unlock the fixing knob
⑤
, move the microscope to
the desired height and lock the fixing knob
⑤
after setting the
height of the microscope. (Fig. 10)
• In order to prevent accidental lowering of the system
due to imperfect locking of the fixing knob, it is advis
-
able to also use the drop prevention ring
⑥
.
9. Use of the microscope
F
Fig. 10
ig. 10
⑤
⑥
①
Summary of Contents for INSPECTION SYSTEMS Series
Page 1: ...Ver 2 1 2021 INSTRUCTION MANUAL INSPECTION SYSTEMS Series Model IS 01 IS 01SMD...
Page 14: ...Page 14 10 Use of the software 10 1 Menu Interface a Control...
Page 15: ...Page 15 b Measurement...
Page 33: ...Modello IS 01 IS 01SMD Serie INSPECTION SYSTEMS MANUALE DI ISTRUZIONI Ver 2 1 2021...
Page 46: ...Pagina 46 10 Uso del software 10 1 Interfaccia del Menu a Controllo...
Page 47: ...Pagina 47 b Misurazioni...
Page 65: ...Ver 2 1 2021 MANUAL DE INSTRUCCIONES Serie INSPECTION SYSTEMS Modelo IS 01 IS 01SMD...
Page 78: ...P gina 78 10 Uso del software 10 1 Interfaz del men a Control...
Page 79: ...P gina 79 b Medici n...
Page 97: ...Ver 2 1 2021 MANUEL D UTILISATION S rie INSPECTION SYSTEMS Mod le IS 01 IS 01SMD...
Page 110: ...Page 110 10 Utilisation du logiciel 10 1 Interface du menu a Contr le...
Page 111: ...Page 111 b Mesures...
Page 129: ...Ver 2 1 2021 BEDIENUNGSANLEITUNG INSPECTION SYSTEMS Series Modell IS 01 IS 01SMD...
Page 142: ...Seite 142 10 Verwendung der Software 10 1 Men Schnittstelle a Kontrolle...
Page 143: ...Seite 143 b Messung...
Page 161: ...Ver 2 1 2021 MANUAL DE INSTRU ES S rie INSPECTION SYSTEMS Modelo IS 01 IS 01SMD...
Page 174: ...P gina 174 10 Uso do software 10 1 Interface do menu a Controlo...
Page 175: ...P gina 175 b Medi o...