background image

Pagina  40

7.2 IS-01SMD

MOUSE 

WIRELESS

DISPOSITIVO 3D

STATI-

COLONNA

MANOPOLA DI 

MESSA A FUOCO

ANELLO DI 

PREVENZIONE 

DISCESA

MANOPOLA ZOOM

MANOPOLA REGOLAZIONE 

INTENSITÀ LUMINOSA

ILLUMINATORE 

ANULARE

CORPO 

MICROSCOPIO

Summary of Contents for INSPECTION SYSTEMS Series

Page 1: ...Ver 2 1 2021 INSTRUCTION MANUAL INSPECTION SYSTEMS Series Model IS 01 IS 01SMD...

Page 2: ...ol 18 10 1 2 Image capture 20 a USB flash drive insertion 20 b Image capture operation 20 c Prefix 20 d Image Format 20 e Image Naming Rules 21 10 1 3 Recording 21 10 1 4 Open Image 22 10 1 5 Image co...

Page 3: ...ty deriving from incorrect instrument use uses that does not comply with this manual 2 Safety Information Avoiding Electrical Shock Before plugging in the power supply make sure that the supplying vol...

Page 4: ...Page 4 Stand Drop prevention ring Focus system Microscope body Ringlight HDMI cable Power supply Wireless mouse Dust cover 3 Package content 3 1 IS 01...

Page 5: ...Page 5 Stand Drop prevention ring Focus system Microscope body Ringlight 3D attachment HDMI cable Power supply Wireless mouse Dust cover 3 2 IS 01SMD...

Page 6: ...glasses Traces of grease or other residuals may deteriorate the final image quality and corrode the optics surface in a short time 5 Intended use Standard models For research and teaching use only Not...

Page 7: ...Page 7 PILLAR FOCUS KNOB STAND DROP PREVENTION RING WIRELESS MOUSE MICROSCOPE BODY 7 Instrument description 7 1 IS 01 ZOOM KNOB RINGLIGHT LIGHT INTENSITY ADJUSTMENT KNOB...

Page 8: ...Page 8 7 2 IS 01SMD PILLAR FOCUS KNOB DROP PREVENTION RING WIRELESS MOUSE MICROSCOPE BODY ZOOM KNOB RINGLIGHT 3D ATTACHMENT LIGHT INTENSITY ADJUSTMENT KNOB STAND...

Page 9: ...Page 9 Back side USB PORT FOR FLASH DRIVE HDMI CABLE CONNECTOR USB PORT FOR MOUSE POWER SUPPLY CONNECTOR...

Page 10: ...1 2 Insert the focus system in the pillar and lock it with fixing knob Fig 2 3 Insert the microscope body in the hole of the focus system and lock it with the fixing knob Fig 3 4 Align the pins of the...

Page 11: ...ort the HDMI cable in the port and the power supply jack in the connector Fig 6 6 Connect a USB flash disk into the port There is no ON OFF switch on the microscope when the power supply is connected...

Page 12: ...e 9 3 Parfocality adjustment 1 Rotate the zoom adjustment ring to the minimum magni fication Fig 9 2 While observing on the monitor focus the specimen 3 Rotate the zoom adjustment ring to the maximum...

Page 13: ...ion that the standard magnification positions are obtained Only for IS 01SMD By sliding the 3D attachment the system allows a quick switching between Straight view and 3D view Fig 11 Once inserted int...

Page 14: ...Page 14 10 Use of the software 10 1 Menu Interface a Control...

Page 15: ...Page 15 b Measurement...

Page 16: ...orner of the menu to switch the menu display status When Locked always show menu When Unlocked the menu will be hidden shown when the mouse is moved to the top left corner of the monitor c Exposure Ex...

Page 17: ...Sharp Saturation Ezoom Contrast adjustable value 0 15 Default value 0 Sharp adjustable value 0 15 Default value 4 Saturation adjustable value 0 254 Default value 128 Ezoom adjustable value 1 6 Defaul...

Page 18: ...rag the crosshair Note Do not press and hold at the intersection of the horizontal and vertical lines Double click a single crosshair to set its color and position center and delete Cross Mark 1 Click...

Page 19: ...icon of the successful insertion of the USB flash drive is displayed for a few seconds in the upper right corner of the display 3 Now it is possible to take an image b Image capture operation 1 Click...

Page 20: ...uring recording the icon in the lower right corner of the screen flashes 5 Click on Record icon again 6 The dialog box Capture Stop appears 7 Click OK to stop recording Note The recorded videos are na...

Page 21: ...Click Full Screen The image interface displays the whole image Click 1 1 The image interface displays part of the picture Press and hold the left mouse button to move the picture position Click Brows...

Page 22: ...ge Comparison image 5 Click Compare Picture icon Image comparison function closes Note Only images in bmp format are supported for image comparison 10 1 6 Freeze 1 Click Freeze icon 2 The current imag...

Page 23: ...e of the current 10 1 9 HDR High Dynamic Range HDR adjustable value 0 3 Default value 1 10 1 10 Anti screen Flicker Anti screen flicker effect is achieved after ticking 10 1 11 Black and White Mode Ch...

Page 24: ...alog box appears In the Line field user can set width and color of the line In the Text field user can set the font size and color of the label b Edge Detection 1 Click Edge Detection icon 2 When sele...

Page 25: ...create a new template 4 After entering a name in the dialog box click Save icon to save the template Click Delete icon to delete this template Click Edit icon to modify the template Once modification...

Page 26: ...select the appropriate calibration and unit 1 Select the appropriate tool for measurement when measuring user only needs to click the points of the object to be measured 2 After you click and release...

Page 27: ...a in the measurement data list with the left mouse button 2 The selected graph data is displayed in shades of gray The color of the selected graph in the image display area becomes the reverse of the...

Page 28: ...t mouse button to pop up the detailed data dialog box to view all the current detailed measurement data h Mark Tool 1 Click Text icon 2 Draw the position to be marked in the image area enter the comme...

Page 29: ...the equipment and that the lamp switch is in off position Users should observe all safety regulations of the region The equipment has acquired the CE safety label However users do have full responsibi...

Page 30: ...t dust on the lenses Clean the lenses The image is not shown on the screen Power supply is unplugged Connect II Mechanical Section Coarse focus knob is hard to turn Tension adjustment ring is too tigh...

Page 31: ...te The separate collection of this equipment at the end of its lifetime is organized and managed by the producer The user will have to contact the manufacturer and follow the rules that he adopted for...

Page 32: ...92 info optikamicroscopes com www optikamicroscopes com OPTIKA Spain spain optikamicroscopes com OPTIKA USA usa optikamicroscopes com OPTIKA China china optikamicroscopes com OPTIKA India india optika...

Page 33: ...Modello IS 01 IS 01SMD Serie INSPECTION SYSTEMS MANUALE DI ISTRUZIONI Ver 2 1 2021...

Page 34: ...tta USB 51 b Operazioni di cattura immagine 51 c Prefisso 51 d Formato Immagine 51 e Regole di denominazione delle immagini 52 10 1 3 Registrazione 52 10 1 4 Apri Immagine 53 10 1 5 Confronto tra imma...

Page 35: ...t derivante da un utilizzo dello strumento non indicato nel presente manuale 2 Informazioni sulla sicurezza Per evitare shock elettrici Prima di collegare il cavo di alimentazione alla presa elettrica...

Page 36: ...36 3 Contenuto della confezione 3 1 IS 01 Stativo Anello prevenzione discesa Sistema di messa a fuoco Corpo del microscopio Illuminatore anulare Cavo HDMI Alimentatore Mouse wireless Copertina antipo...

Page 37: ...Pagina 37 3 2 IS 01SMD Stativo Anello prevenzione discesa Sistema di messa a fuoco Corpo del microscopio Illuminatore anulare Dispositivo 3D Cavo HDMI Alimentatore Mouse wireless Copertina antipolvere...

Page 38: ...che come lenti filtri o vetri Tracce di grasso o altri residui possono ridurre la qualit visiva dell immagine finale e corrodere la superficie delle ottiche in breve tempo 5 Utilizzo previsto Modelli...

Page 39: ...scrizione dello strumento 7 1 IS 01 COLONNA MANOPOLA DI MESSA A FUOCO STATIVO ANELLO DI PREVENZIONE DISCESA MOUSE WIRELESS CORPO MICROSCOPIO MANOPOLA ZOOM ILLUMINATORE ANULARE MANOPOLA REGOLAZIONE INT...

Page 40: ...40 7 2 IS 01SMD MOUSE WIRELESS DISPOSITIVO 3D STATI COLONNA MANOPOLA DI MESSA A FUOCO ANELLO DI PREVENZIONE DISCESA MANOPOLA ZOOM MANOPOLA REGOLAZIONE INTENSIT LUMINOSA ILLUMINATORE ANULARE CORPO MICR...

Page 41: ...Pagina 41 Lato posteriore PORTA USB PER CHIAVE USB CONNETTORE CAVO HDMI PORTA USB PER MOUSE CONNETTORE ALIMENTAZIONE...

Page 42: ...con la manopola di fissaggio Fig 2 3 Inserire il corpo del microscopio nel foro del sistema di messa a fuoco e bloccarlo con la manopola di fissaggio Fig 3 4 Allineare i pin del corpo del microscopio...

Page 43: ...o HDMI nella porta e lo spinotto dell alimentatore nella porta Fig 6 6 Collegare una chiave USB nella porta Non c un interruttore ON OFF sul microscopio quando l alimentazione collegata il sistema si...

Page 44: ...andimento Ora la regolazione della parfocalit dovrebbe essere impostata 9 4 Messa a fuoco Mettere il campione da osservare sul piatto del supporto e met tere a fuoco il campione usando le manopole di...

Page 45: ...tenere un osservazione obliqua dell intero campione 1 Mentre si guarda sul monitor ruotare il dispositivo 3D fino ad ottenere l immagine desiderata 9 5 Variazione dell ingrandimento 1 Ruotare l anello...

Page 46: ...Pagina 46 10 Uso del software 10 1 Interfaccia del Menu a Controllo...

Page 47: ...Pagina 47 b Misurazioni...

Page 48: ...per cambiare lo stato di visualizzazione del menu Quando Chiuso mostra sempre il menu Quando Aperto il menu sar nascosto mostrato quando il mouse viene spostato nell angolo in alto a sinistra del moni...

Page 49: ...ntrast Sharp Saturation Ezoom Contrast valore regolabile 0 15 Valore di default 0 Sharp valore regolabile 0 15 Valore di default 4 Saturation valore regolabile 0 254 Valore di defaulte 128 Ezoom valor...

Page 50: ...re il crocefilo Nota non tenere premuto all intersezione delle linee orizzontali e verticali Fare doppio clic su un singolo crocefilo per impostarne il colore e la posizione centrare e cancellare Marc...

Page 51: ...nserito la chiavetta USB l icona dell avvenuto inserimento della chiavetta USB viene visualizzata per alcuni secondi nell angolo superiore destro del monitor 3 Ora possibile acquisire un immagine b Op...

Page 52: ...e 4 Durante la registrazione l icona nell angolo in basso a destra dello monitor lampeggia 5 Cliccare nuovamente l icona Record 6 Appare la finestra di dialogo Capture Stop 7 Cliccare OK per interromp...

Page 53: ...rfaccia immagine visualizza l intera immagine Cliccare su 1 1 L interfaccia immagine visualizza una parte dell immagine Tenere premuto il tasto sinistro del mouse per spostare la posizione dell immagi...

Page 54: ...liccare sull icona Compare Picture La funzione di confronto delle immagini si chiude Nota solo le immagini in formato bmp sono supportate per il confronto delle immagini 10 1 6 Congela 1 Cliccare sull...

Page 55: ...HDR valore regolabile 0 3 valore di default 1 10 1 10 Antisfarfallio dello schermo L effetto di eliminazione dello sfarfallio dello schermo si ottiene dopo l abilitazione 10 1 11 Modalit Bianco Nero A...

Page 56: ...o delle impostazioni Nel campo Line l utente pu impostare la larghezza e il colore della linea Nel campo Text l utente pu impostare la dimensione del carattere e il colore dell etichetta b Rilevazione...

Page 57: ...er inserito un nome nella finestra di dialogo cliccare sull icona Save per salvare il template Cliccare sull icona Delete per eliminare questo template Cliccare sull icona Edit per modificare questo t...

Page 58: ...selezionare la calibrazione e l unit appropriate 1 Selezionare lo strumento appropriato per la misurazione quando si misura l utente deve solo cliccare i punti dell oggetto da misurare 2 Dopo aver cl...

Page 59: ...azione con il tasto sinistro del mouse 2 I dati della grafica selezionata vengono visualizzati in tonalit di grigio Il colore della grafica selezionata nell area di visualizzazione dell immagine diven...

Page 60: ...finestra di dialogo dei dati dettagliati per visualizzare tutti i dati di misurazione dettagliati correnti h Strumento Segna 1 Cliccare sull icona Text 2 Disegna la posizione da marcare nell area dell...

Page 61: ...izionato su off Attenersi a tutte le precauzioni di sicurezza della zona in cui ci si trova ad operare L apparecchio omologato secondo le norme di sicurezza CE Gli utenti hanno comunque piena responsa...

Page 62: ...L immagine non appare sullo schermo I connettori dell alimentatore non sono ben collegati Collegarli II Sezione Meccanica La manopola macrometrica difficile da ruotare L anello di regolazione della te...

Page 63: ...ne vita organizzata e gestita dal produttore L utente che vorr disfarsi della presente apparecchiatura dovr quindi contattare il produttore e seguire il sistema che questo ha adottato per consentire l...

Page 64: ...92 info optikamicroscopes com www optikamicroscopes com OPTIKA Spain spain optikamicroscopes com OPTIKA USA usa optikamicroscopes com OPTIKA China china optikamicroscopes com OPTIKA India india optika...

Page 65: ...Ver 2 1 2021 MANUAL DE INSTRUCCIONES Serie INSPECTION SYSTEMS Modelo IS 01 IS 01SMD...

Page 66: ...Operaci n de captura de im genes 83 c Prefijo 83 d Formato de la imagen 83 e Reglas de denominaci n de las im genes 84 10 1 3 Grabaci n 84 10 1 4 Abrir la imagen 85 10 1 5 Comparaci n de im genes 86...

Page 67: ...ellos que utilizan dicho instrumento 2 Informaci n de seguridad Evitar una descarga el ctrica Antes de conectar el microscopio a la toma de corriente asegurarse que la tensi n de entrada del lugar don...

Page 68: ...gina 68 Soporte Anillo de prevenci n de ca da Sistema de enfoque Cuerpo del microscopio Iluminador anular Cable HDMI Alimentaci n Rat n inal mbrico Cubierta antipolvo 3 Contenido del paquete 3 1 IS 0...

Page 69: ...P gina 69 Soporte Anillo de prevenci n de ca da Sistema de enfoque Cuerpo del microscopio Iluminador anular Dispositivo 3D Cable HDMI Alimentaci n Rat n inal mbrico Cubierta antipolvo 3 2 IS 01SMD...

Page 70: ...tros o el cristal Los restos de grasa u otros residuos pueden reducir la calidad visual de la imagen final y corroer la superficie de la ptica en poco tiempo 5 Utilizaci n Modelos est ndar Para uso ex...

Page 71: ...LAR SOPORTE ANILLO DE PREVENCI N DE CA DA RAT N INAL MBRICO CUERPO DEL MICROSCOPIO 7 Descripci n del instrumento 7 1 IS 01 MANDO AJUSTE DE LA INTENSIDAD DE LA LUZ MANDO DE ENFOQUE MANDO DEL ZOOM ILUMI...

Page 72: ...72 7 2 IS 01SMD PILAR MANDO DE ENFOQUE ANILLO DE PREVENCI N DE CA DA RAT N INAL MBRICO CUERPO DEL MICROSCOPIO MANDO DEL ZOOM ILUMINADOR ANULAR DISPOSITIVO 3D MANDO AJUSTE DE LA INTENSIDAD DE LA LUZ S...

Page 73: ...P gina 73 Parte posterior PUERTO USB PARA MEMORIA USB CONECTOR CABLE HDMI PUERTO USB PARA EL RAT N CONECTOR ALIMENTACI N...

Page 74: ...tema de enfoque en el pilar y bloquearlo con la perilla de fijaci n Fig 2 3 Insertar el cuerpo del microscopio en el orificio del sistema de enfoque y bloquearlo con la perilla de fijaci n Fig 3 4 Ali...

Page 75: ...erto USB el cable HDMI en el puerto y la toma de alimentaci n en el conector Fig 6 6 Conectar una memoria USB el puerto El microscopio no tiene un interruptor ON OFF cuando se conecta la fuente de ali...

Page 76: ...que el enfoque sea estable 9 3 Ajuste de la parfocalidad 1 Girar el anillo de ajuste del zoom hasta el m nimo aumen to Fig 9 2 Mientras se observa en el monitor enfocar la muestra 3 Girar el anillo d...

Page 77: ...n de que se obtienen las posiciones de aumento est ndar S lo para IS 01SMD Al deslizar el accesorio 3D el sistema permite cambiar r pidamente entre la vista recta y la vista 3D Fig 11 Una vez insertad...

Page 78: ...P gina 78 10 Uso del software 10 1 Interfaz del men a Control...

Page 79: ...P gina 79 b Medici n...

Page 80: ...ualizaci n del men Cuando est Bloqueado siempre muestra el men Cuando est Desbloqueado el men se ocultar mostrar cuando el rat n se mueva a la esquina superior izquierda del monitor c Exposici n El aj...

Page 81: ...arp Saturation Ezoom Contrasto valor ajustable 0 15 Valor por defecto 0 Nitidez valor ajustable 0 15 Valor por defecto 4 Saturaci n valor ajustable 0 254 Valor por defecto 128 Ezoom valor ajustable 1...

Page 82: ...arrastrar el ret culo Nota No mantener pulsado en la intersecci n de las l neas horizontales y verticales Hacer doble clic en una sola cruz para fijar su color y posici n centrarla y eliminarla Marca...

Page 83: ...la memoria USB el icono de la inserci n exitosa de la memoria USB se muestra durante unos segundos en la esquina superior derecha de la pantalla 3 Ahora es posible capturar una imagen b Operaci n de...

Page 84: ...grabaci n 4 Durante la grabaci n el icono de la esquina inferior derecha de la pantalla parpadea 5 Pulsar el icono Record de nuevo 6 Aparece el cuadro de di logo Capture Stop 7 Pulsar OK para detener...

Page 85: ...imagen muestra la imagen completa Pulsar 1 1 La interfaz de la imagen muestra una parte de la misma Mantener pulsado el bot n izquierdo del rat n para mover la posici n de la imagen Pulsar Browse par...

Page 86: ...comparaci n 5 Pulsar el icono Compare Picture La funci n de comparaci n de im genes se cierra Nota S lo se admiten im genes en formato bmp para la comparaci n de im genes 10 1 6 Congelar 1 Pulsar el...

Page 87: ...h Dynamic Range HDR valor 0 3 ajustable Valor por defecto 1 10 1 10 Antiparpadeo de la pantalla El efecto antiparpadeo de la pantalla se consigue tras la marcadura 10 1 11 Modo blanco y negro Marcar p...

Page 88: ...uraci n En el campo Line el usuario puede establecer el ancho y el color de la l nea En el campo Text el usuario puede establecer el tama o de la letra y el color de la etiqueta b Detecci n de Bordes...

Page 89: ...u s de introducir un nombre en el cuadro de di logo pulsar el icono Save para guardar el modelo Pulsar el icono Delete para eliminar este modelo Pulsar el icono Edit para modificar el modelo Una vez h...

Page 90: ...ebe seleccionar primero la calibraci n y la unidad adecuadas 1 Seleccionar la herramienta adecuada para la medici n al medir el usuario s lo necesita pulsar los puntos del objeto a medir 2 Despu s de...

Page 91: ...n con el bot n izquierdo del rat n 2 Los datos del gr fico seleccionado se muestran en tonos de gris El color del gr fico seleccionado en el rea de visualizaci n de la imagen se convierte en el invers...

Page 92: ...parezca el cuadro de di logo de datos detallados para ver todos los datos de medici n detallados actuales h Herramienta de Marcado 1 Pulsar el icono Text 2 Dibujar la posici n a marcar en el rea de la...

Page 93: ...n del microscopio y que el interruptor del iluminador est en la posici n off El usuario debe consultar las normas de seguridad de su pa s El instrumento est dotado de una etiqueta de seguridad CE No o...

Page 94: ...gen no aparece en la pantalla El enchufe no est conectado al siste ma de iluminaci n Conectar II Secci n Mec nica El mando macrom trico gira con difi cultad El anillo de regulaci n de la tensi n est d...

Page 95: ...organizada y gestionada por el fabricante Por lo tanto el usuario que desee deshacerse de este equipo debe ponerse en contacto con el fabricante y seguir el sistema que ha adoptado para permitir la r...

Page 96: ...92 info optikamicroscopes com www optikamicroscopes com OPTIKA Spain spain optikamicroscopes com OPTIKA USA usa optikamicroscopes com OPTIKA China china optikamicroscopes com OPTIKA India india optika...

Page 97: ...Ver 2 1 2021 MANUEL D UTILISATION S rie INSPECTION SYSTEMS Mod le IS 01 IS 01SMD...

Page 98: ...l USB 115 b Op ration de capture d image 115 c Pr fixe 115 d Format de l image 115 e R gles de d nomination des images 116 10 1 3 Enregistrement 116 10 1 4 Ouvrir une image 117 10 1 5 Comparaison d im...

Page 99: ...t instrument Nous d clinons toute responsabilit quant des utilisations de l instrument non conformes au pr sent manuel 2 Pr cautions viter choc lectrique Avant de connecter le c ble d alimentation au...

Page 100: ...ort Bague de pr vention des chutes Syst me de mise au point Corps du microscope lluminateur annulaire C ble HDMI Alimentation lectrique Souris sans fil Couverture anti poussi re 3 Contenu de l emballa...

Page 101: ...Page 101 Support Bague de pr vention des chutes Syst me de mise au point Corps du microscope lluminateur annulaire 3D attachment HDMI cable Power supply Wireless mouse Dust cover 3 2 IS 01SMD...

Page 102: ...ts de l huile de graisse ou d autres r sidus sur les objectifs les filtres les verres diminuent g n ralement la clarit d image 5 Emploi pr vu Mod les standard R serv la recherche et l enseignement Ne...

Page 103: ...ILIER BOUTON DE MISE AU POINT SUPPORT BAGUE ANTI CHUTE SOURIS SANS FIL CORPS DU MICROSCOPE 7 Description de l instrument 7 1 IS 01 BOUTON DU ZOOM ILLUMINATEUR ANNULAIRE BOUTON DE R GLAGE DE L INTENSIT...

Page 104: ...104 7 2 IS 01SMD ATTACHEMENT 3D PILIER BOUTON DE MISE AU POINT SUPPORT BAGUE ANTI CHUTE SOURIS SANS FIL CORPS DU MICROSCOPE BOUTON DU ZOOM BOUTON DE R GLAGE DE L INTENSIT LUMINEUSE ILLUMINATEUR ANNUL...

Page 105: ...Page 105 C t arri re PORT USB POUR CL USB CONNECTEUR DE C BLE HDMI PORT USB POUR SOURIS CONNECTEUR D ALIMENTATION...

Page 106: ...t dans le pilier et verrouillez le avec le bouton de fixation Fig 2 3 Ins rez le corps du microscope dans le trou du syst me de mise au point et verrouillez le avec le bouton de fixation Fig 3 4 Align...

Page 107: ...ns le port USB le c ble HDMI dans le port et la prise d alimentation dans le connecteur Fig 6 6 Connectez une cl USB dans le port Il n y a pas d interrupteur ON OFF sur le microscope lorsque l aliment...

Page 108: ...stable 9 3 Ajustement de la parfocalit 1 Tournez la bague de r glage du zoom jusqu au grossis sement minimum Fig 9 2 Tout en observant sur le moniteur mettez l chantillon au point 3 Tournez la bague d...

Page 109: ...es positions de grossissement standard sont obtenues Uniquement pou IS 01SMD En faisant glisser l attachement 3D le syst me permet de passer rapidement de la vue droite la vue 3D Fig 11 Une fois ins r...

Page 110: ...Page 110 10 Utilisation du logiciel 10 1 Interface du menu a Contr le...

Page 111: ...Page 111 b Mesures...

Page 112: ...l tat d affichage du menu Lorsqu il est Verrouill il affiche toujours le menu Lorsqu il est D verrouill le menu sera cach affich lorsque la souris sera d plac e vers le coin sup rieur gauche de l cran...

Page 113: ...harp Saturation Ezoom Contrast valeur r glable de 0 15 Valeur par d faut 0 Sharp valeur r glable de 0 15 Valeur par d faut 4 Saturation valeur r glable de 0 254 Valeur par d faut 128 Ezoom valeur r gl...

Page 114: ...r ticule Remarque Ne pas appuyer et maintenir l intersection des lignes horizontales et verticales Double Cliquer sur un seul r ticule pour d finir sa couleur et sa position le centrer et le supprime...

Page 115: ...a cl USB l ic ne de l insertion r ussie de la cl USB s affiche pendant quelques secondes dans le coin sup rieur droit de l cran 3 Il est d sormais possible de prendre une image b Op ration de capture...

Page 116: ...Pendant l enregistrement l ic ne dans le coin inf rieur droit de l cran clignote 5 Cliquez nouveau sur l ic ne Record 6 La bo te de dialogue Capture Stop appara t 7 Cliquer OK pour arr ter l enregist...

Page 117: ...een L interface image affiche l image enti re Cliquer 1 1 L interface image affiche une partie de l image Appuyez et maintenez le bouton gauche de la souris pour d placer la position de l image Clique...

Page 118: ...ence 5 Cliquer sur l ic n Compare Picture La fonction de comparaison d images se ferme Remarque seules les images au format bmp sont prises en charge pour la comparaison d images 10 1 6 Geler 1 Clique...

Page 119: ...1 9 HDR High Dynamic Range HDR valeur r glable de 0 3 Valeur par d faut 1 10 1 10 Anti scintillement d cran L effet anti scintillement de l cran est obtenu apr s le cochage 10 1 11 Mode Noir et Blanc...

Page 120: ...s le champ Line l utilisateur peut d finir la largeur et la couleur de la ligne Dans le champ Text l utilisateur peut d finir la taille de la lettre et la couleur de l tiquette b D tection des Bords 1...

Page 121: ...od le 4 Apr s avoir saisi un nom dans la bo te de dialogue cliquer sur l ic ne Save pour sauver le mod le Cliquer sur l ic ne Delete pour supprimer ce mod le Cliquer sur l ic ne Edit pour modifier le...

Page 122: ...brage et l unit appropri s 1 S lectionner l outil appropri pour la mesure lors de la mesure l utilisateur doit seulement cliquer sur les points de l objet mesurer 2 Apr s avoir cliqu et rel ch la sour...

Page 123: ...es de mesure avec le bouton gauche de la souris 2 Les donn es du graphique s lectionn sont affich es en nuances de gris La couleur du graphique s lectionn dans la zone d affichage de l image devient...

Page 124: ...a bo te de dialogue des donn es d taill es afin de visualiser toutes les donn es de mesure d taill es actuelles h Outil de Marquage 1 Cliquer sur l ic ne Text 2 Dessinez la position marquer dans la zo...

Page 125: ...e l clairage soit en position arr t L utilisateur devra consulter les normes de s curit s de son pays L appareil incl t une tiquette de s curit C E Dans tous les cas l utilisateur assume toute respon...

Page 126: ...llon Salet poussi re sur les lentilles Nettoyer les lentilles L image n appara t pas l cran L alimentation lectrique est d bran ch e Brancher le correctement II Section M canique Commande macrom triqu...

Page 127: ...du pr sent instrument sera effectu e par le fabricant Par cons quent l utilisateur qui souhaite liminer l appareil devra se mettre en contact avec le fabricant et suivre le syst me que celui ci a adop...

Page 128: ...92 info optikamicroscopes com www optikamicroscopes com OPTIKA Spain spain optikamicroscopes com OPTIKA USA usa optikamicroscopes com OPTIKA China china optikamicroscopes com OPTIKA India india optika...

Page 129: ...Ver 2 1 2021 BEDIENUNGSANLEITUNG INSPECTION SYSTEMS Series Modell IS 01 IS 01SMD...

Page 130: ...145 10 1 2 Bildaufnahme 147 a Einsetzen eines USB Sticks 147 b Bildaufnahmevorgang 147 c Pr fix 147 d Bildformat 147 e Regeln f r die Bildbenennung 148 10 1 3 Aufnahme 148 10 1 4 Bild ffnen 149 10 1 5...

Page 131: ...n Benutzern zur Verf gung stehen Wir lehnen jede Verantwortung f r eine fehlerhafte in dieser Bedienungsanleitung nicht gezeigten Verwendung Ihrer Produkte ab 2 Sicherheitsinformationen Elektrische En...

Page 132: ...Seite 132 Stand Fallschutzring Fokus System Mikroskopk rper Ringlicht HDMI Kabel Netzteile Kabellose Maus Staubschutzh lle 3 Verpackungsinhalt 3 1 IS 01...

Page 133: ...Seite 133 Stand Fallschutzring Fokus System Mikroskopk rper Ringlicht 3D Anlage HDMI Kabel Netzteile Kabellose Maus Staubschutzh lle 3 2 IS 01SMD...

Page 134: ...anderen R ckst nden k nnen die endg ltige Bildqualit t beeintr chtigen und die Optikoberfl che in kurzer Zeit angreifen 5 Verwendung Standardmodelle Nur f r Forschung und Lehre verwenden Nicht f r th...

Page 135: ...Seite 135 S ULE FOKUSKNOPF STAND FALLSCHUTZRING KABELLOSE MAUS MIKROSKOPK RPER 7 Beschreibung des Instruments 7 1 IS 01 ZOOM KNOPF RINGLICHT EINSTELLKNOPF F R DIE LICHTST RKE...

Page 136: ...Seite 136 7 2 IS 01SMD 3D ANLAGE STAND S ULE FOKUSKNOPF FALLSCHUTZRING KABELLOSE MAUS MIKROSKOPK RPER ZOOM KNOPF RINGLICHT EINSTELLKNOPF F R DIE LICHTST RKE...

Page 137: ...Seite 137 R ckseite USB ANSCHLUSS F R USB STICK HDMI KABE LANSCHLUSS USB ANSCHLUSS F R MAUS STROMVERSOR GUNGSSTECKER...

Page 138: ...en das Fokussystem in die S ule ein und verriegeln es mit dem Befestigungsknopf Fig 2 3 Setzen das Mikroskopk rpers in die ffnung des Fokussystems ein und verriegeln Sie es mit dem Feststellknopf Fig...

Page 139: ...I Kabel an den Anschluss und die Stromversorgungsbuchse an den Anschluss an Fig 6 6 Schlie en Sie einen USB Stick an den Anschluss an Am Mikroskop gibt es keinen ON OFF Schalter Sobald die Stromversor...

Page 140: ...ellung der Parfokalit t 1 Drehen den Zoom Einstellring auf die minimale Vergr erung Fig 9 2 Fokussieren das Probe w hrend es auf dem Monitor be obachten 3 Drehen den Zoom Einstellring auf die maximale...

Page 141: ...dass die Standardvergr erungspositionen erreicht sind Nur f r IS 01SMD Durch Verschieben des 3D Anlage erm glicht das System einen schnellen Wechsel zwischen gerader Ansicht und 3D Ansicht Fig 11 Nac...

Page 142: ...Seite 142 10 Verwendung der Software 10 1 Men Schnittstelle a Kontrolle...

Page 143: ...Seite 143 b Messung...

Page 144: ...en anzeige zu ndern Wenn Gesperrt immer Men anzeigen Wenn Entsperrt wird das Men ein ausgeblendet wenn die Maus in die linke obere Ecke des Monitors bewegt wird c Exposition Die Belichtung kann Auto o...

Page 145: ...om eingestellt werden Contrast einstellbarer Wert 0 15 Standardwert 0 0 Sharp einstellbarer Wert 0 15 Standardwert 4 Saturation einstellbarer Wert 0 254 Standardwert 128 Ezoom einstellbarer Wert 1 6 S...

Page 146: ...en Hinweis Dr cken und halten nicht am Schnittpunkt der horizontalen und vertikalen Linien Doppelklicken auf ein einzelnes Fadenkreuz um seine Farbe und Position festzulegen zu zentrieren und zu l sch...

Page 147: ...B Sticks wird das Symbol f r das erfolgreiche Einstecken des USB Sticks f r einige Sekunden in der oberen rechten Ecke des Displays angezeigt 3 Jetzt ist es m glich ein Bild aufzunehme b Bildaufnahmev...

Page 148: ...der Aufnahme blinkt das Symbol in der unteren rechten Ecke des Bildschirms 5 Klicken erneut auf das Symbol Record 6 Das Dialogfenster Capture Stop erscheint 7 Klicken auf OK um die Aufnahme zu beende...

Page 149: ...ll Screen Die Bildschnittstelle zeigt das gesamte Bild an Klicken auf 1 1 Die Bildschnittstelle zeigt einen Teil des Bildes an Halten die linke Maustaste gedr ckt um die Bildposition zu verschieben Kl...

Page 150: ...eichsbild 5 Klicken auf das Symbol Compare Picture Bildvergleichsfunktion wird geschlossen Hinweis Es werden nur Bilder im bmp Format f r den Bildvergleich unterst tzt 10 1 6 Gefrieren 1 Klicken auf d...

Page 151: ...uellen Richtung 10 1 9 HDR High Dynamic Range HDR einstellbarer Wert 0 3 Standardwert 1 10 1 10 Anti Bildschirmflimmern Anti Bildschirm Flackereffekt wird nach dem Ticken erreicht 10 1 11 Schwarz Wei...

Page 152: ...eints Im Feld Line kann der Benutzer Breite und Farbe der Linie einstellen Im Feld Text kann der Benutzer die und die Farbe des Etiketts einstellen b Kantenerkennung 1 Klicken auf das Symbol Edge Dete...

Page 153: ...n 4 Nach der Eingabe eines Namens im Dialogfeld klicken Save um die Vorlage zu speichern Klicken auf das Symbol Delete um diese Vorlage zu l schen Klicken auf das Symbol Edit um die Vorlage zu ndern N...

Page 154: ...rechende Kalibrierung und Einheit ausw hlent 1 Auswahl des geeigneten Werkzeugs f r die Messung beim Messen muss der Benutzer nur die Punkte des zu messenden Objekts anklicken 2 Nachdem Sie die Maus a...

Page 155: ...f die Daten in der Messdatenliste 2 Die ausgew hlten Diagrammdaten werden in Graut nen angezeigt Die Farbe der ausgew hlten Grafik im Bildanzeigebereich wird umgekehrt zur eingestellten Farbe f Grafik...

Page 156: ...Dialogfeld Detaildaten zu ffnen in dem alle aktuellen detaillierten Messdaten angezeigt werden h Werkzeug markieren 1 Klicken auf das Symbol Text 2 Zeichnen die zu markierende Stelle im Bildbereich ge...

Page 157: ...e Spannung f r das Mikroskop geeignet ist und dass der Beleuchtungsschalter sich in position OFF befindet Beachten Sie alle Sicherheitsvorschriften des Arbeitsplatzes an dem Sie mit dem Mikroskop arbe...

Page 158: ...Reinigen die Linsen Das Bild wird nicht auf dem Bildschirm angezeigt Stromversorgungsstecker sind nicht gut angeschlossen Verbinden sie II Mechanischer System Der makrometrische Knopf ist schwer zu dr...

Page 159: ...ennte Sammlung von Ger ten die am Ende Ihrer Lebensdauer sind wird vom Hersteller organisiert Der Benutzer der dieses Ger t entsorgen m chtet muss dann Kontakt mit dem Hersteller aufnehmen und der Vor...

Page 160: ...92 info optikamicroscopes com www optikamicroscopes com OPTIKA Spain spain optikamicroscopes com OPTIKA USA usa optikamicroscopes com OPTIKA China china optikamicroscopes com OPTIKA India india optika...

Page 161: ...Ver 2 1 2021 MANUAL DE INSTRU ES S rie INSPECTION SYSTEMS Modelo IS 01 IS 01SMD...

Page 162: ...a Inser o de pen drive USB 179 b Opera o de captura de imagem 179 c Prefixo 179 d Formato da imagem 179 e Regras de nomea o de imagens 180 10 1 3 Grava o 180 10 1 4 Abrir imagem 181 10 1 5 Compara o d...

Page 163: ...e qualquer responsabilidade em caso de Uso do instrumento n o indicada neste manual 2 Informa es sobre a seguran a Para evitar choques el ctricos Antes de ligar o cabo de alimenta o com a tomada el ct...

Page 164: ...P gina 164 Stand Anel de preven o de quedas Sistema de focagem Corpo do microsc pioy Iluminador anular Cabo HDMI Alimenta o el ctrica Rato sem fios Capa de poeira 3 Conte do da embalagem 3 1 IS 01...

Page 165: ...P gina 165 Stand Anel de preven o de quedas Sistema de focagem Corpo do microsc pio Iluminador anular Acess rio 3D Cabo HDMI Alimenta o el ctrica Rato sem fios Capa de poeira 3 2 IS 01SMD...

Page 166: ...gios de graxa ou outros res duos podem deteriorar a qualidade final da imagem e corroer a superf cie ptica em pouco tempo 5 Uso previsto Modelos padr o Apenas para uso em pesquisa e ensino N o se des...

Page 167: ...a 167 PILAR BOT O DE FOCAGEM STAND ANEL DE PREVEN O DE QUEDAS RATO SEM FIOS CORPO DO MICROSC PIO 7 Descri o do instrumento 7 1 IS 01 BOT O DE ZOOM ILUMINADOR ANULAR BOT O DE AJUSTE DE INTENSIDADE LUMI...

Page 168: ...P gina 168 7 2 IS 01SMD ACESS RIO 3D STAND PILAR BOT O DE FOCAGEM ANEL DE PREVEN O DE QUEDAS RATO SEM FIOS CORPO DO MICROSC PIO BOT O DE ZOOM ILUMINADOR ANULAR BOT O DE AJUSTE DE INTENSIDADE LUMINOSA...

Page 169: ...P gina 169 Parte de tr s PORTA USB PARA PEN DRIVE CONECTOR DE CABO HDMI PORTA USB PARA RATO CONECTOR FONTE DE ALIMENTA O...

Page 170: ...r o sistema de focagem no pilar e bloque lo com o bot o de fixa o Fig 2 3 Inserir o corpo do microsc pio no orif cio do sistema de focagem e bloque lo com o bot o de fixa o Fig 3 4 Alinhar os pinos do...

Page 171: ...sem fios na porta USB o cabo HDMI na porta e a tomada da fonte de alimenta o no conector Fig 6 6 Ligar um USB flash disk porta N o h interruptor ON OFF no microsc pio quando a alimenta o ligada o sist...

Page 172: ...mb m tornar a focagem est vel 9 3 Ajuste da parfocalidade 1 Rodar o anel de ajuste de zoom para a amplia o m ni ma Fig 9 2 Enquanto observa no monitor focalize o esp cime 3 Rodar o anel de ajuste de z...

Page 173: ...ndica o de que as posi es de amplia o padr o s o obtidas Apenas para IS 01SMD Ao deslizar o acess rio 3D o sistema permite uma r pida comuta o entre a vista recta e a vista 3D Fig 11 Uma vez inserido...

Page 174: ...P gina 174 10 Uso do software 10 1 Interface do menu a Controlo...

Page 175: ...P gina 175 b Medi o...

Page 176: ...o do menu para mudar o estado de exibi o do menu Quando Bloqueado mostrar sempre o menu Quando Desbloqueado o menu ser escondido mostrado quando o rato movido para o canto superior esquerdo do monitor...

Page 177: ...tion Ezoom Contrast valor ajust vel 0 15 Valor por defeito 0 Sharp valor ajust vel 0 15 Valor por defeito 4 Saturation valor ajust vel 0 254 Valor por defeito 128 Ezoom valor ajust vel 1 6 Valor por d...

Page 178: ...pode arrastar o ret culo Nota N o pressionar e segurar na intersec o das linhas horizontais e verticais Fa a duplo Clicar numa nico reticulo para definir o suo cor e posi o centrar e apagar Reticulo...

Page 179: ...da pen drive USB o cone da inser o bem sucedida da pen drive USB exibido durante alguns segundos no canto superior direito do monitor 3 Agora poss vel ter uma imagem b Opera o de captura de imagem 1...

Page 180: ...niciar a grava o 4 Durante a grava o o cone no canto inferior direito do ecr pisca 5 Clicar novamente no cone Record 6 Aparece a caixa de di logo Capture Stop 7 Clicar em OK para parar de gravar Nota...

Page 181: ...ar Full Screen A interface de imagem mostra a imagem inteira Clicar 1 1 A interface de imagem mostra parte da imagem Manter premido o bot o esquerdo do rato para mover a posi o da image Clicar Browse...

Page 182: ...comparativa 5 Clicar no cone Compare Picture A fun o de compara o de imagens fecha Nota Apenas imagens em formato bmp s o suportadas para compara o de imagens 10 1 6 Congelar 1 Clicar no cone Freeze...

Page 183: ...High Dynamic Range HDR valor ajust vel 0 3 Valor por defeito 1 10 1 10 Anti flicker de ecr O efeito de cintila o do ecr alcan ado ap s o fecho 10 1 11 Modo Preto e Branco Marcar para o modo preto e br...

Page 184: ...ini o No campo Line o utilizador pode definir a largura e a cor da linha No campo Text o utilizador pode definir o tamanho da fonte e a cor da etiqueta b Detec o de bordas 1 Clicar no cone Edge Detect...

Page 185: ...delo 4 Depois de introduzir um nome na caixa de di logo clicar no cone Save para guardar o modelo Clicar no cone Delete para apagar este modelo Clicar no cone Edit para modificar o modelo Uma vez que...

Page 186: ...imeiro seleccionar a calibra o e unidade adequadas 1 Seleccionar a ferramenta apropriada para medi o ao medir o utilizador s precisa de clicar nos pontos do objecto a ser medido 2 Depois de clicar e s...

Page 187: ...de dados de medi o com o bot o esquerdo do rato 2 Os dados gr ficos seleccionados s o apresentados em tons de cinza A cor do gr fico seleccionado na rea de visualiza o da imagem torna se o inverso da...

Page 188: ...o para abrir a caixa de di logo de dados detalhados para ver todos os dados de medi o detalhados actuais h Ferramenta de Marca 1 Clicar no cone Text 2 Desenhe a posi o a ser marcada na rea de imagem i...

Page 189: ...steja posicionado no off Seguir todas as precau es de seguran a da zona na qual se trabalha O aparelho aprovado segundo as normas de seguran a CE Os utilizadores t m de qualquer modo plena responsabil...

Page 190: ...H manchas e p na lentes Limpe a lente A imagem n o mostrada no ecr Conectores de alimenta o n o devi damente ligados Conecte os II Sec o Mec nica O bot o do foco macro est dif cil de rodar O anel de a...

Page 191: ...dos outros res duos A recolha separada deste equipamento no final da sua vida til organizada e gerida pelo produtor O utilizador ter de contactar o fabricante e seguir as regras que adoptou para a re...

Page 192: ...optikamicroscopes com OPTIKA USA usa optikamicroscopes com OPTIKA China china optikamicroscopes com OPTIKA India india optikamicroscopes com OPTIKA Central America camerica optikamicroscopes com 100...

Reviews: