Page 6
•
Fully insert the phase contrast condenser into the holder, pushing it until it is well inserted:
•
Rotate the condenser annuli disc to the “0” position (brightfield).
•
Completely open the aperture diaphragm:
•
Turn on the light and put a sample slide on the stage (an opaque specimen will make alignment easier).
•
Using the 10x objective, center your sample (moving the stage) and focus using the coarse and fine focus
knobs.
•
Rotate the condenser height adjustment knob until you obtain a uniform white illumination on the sample.
Important: the optimal position corresponds to the top lens of the condenser 1-2mm below the bottom side
of the slide.
•
Rotate the condenser annuli disc to the “10” position.
Completely open this diaphragm
Summary of Contents for B-1000 POL
Page 1: ...Version 2 Issued 18 08 2014 Model B 380 B 500 B 800 B 1000 Phase contrast INSTRUCTION MANUAL...
Page 25: ......
Page 26: ......
Page 39: ......
Page 40: ......
Page 41: ...Mod le B 380 B 500 B 800 B 1000 Contrasto di fase MANUEL D UTILISATION Version 2 du 18 08 2014...
Page 65: ...Page 65...
Page 66: ......
Page 67: ...Modell B 380 B 500 B 800 B 1000 Phasenkontrast ANLEITUNG Version 2 Ausgestellt am 18 08 2014...
Page 79: ......