Page 7
Image as seen in the telescope
Levers for phase ring alignment
•
Take one of the eyepieces out of the tube, and insert the centering telescope. Rotate its upper part in order
to view a sharp image of the phase ring:
•
Typically, you have to see a pair of rings: one brighter and one darker. The bright one is the image of the an
-
nulus on the condenser, while the dark one is the phase ring inside the objective:
•
Rotate the levers on both sides of the condenser. You will notice that the bright ring will change its position:
Centering telescope