10
SCALE SETTINGS
To Enter the Settings Mode:
• While pressing [MODE] and [TARE], turn on the power. The scale will
enter the setting mode. Press [MODE] to select different settings which
you can set using the [UNIT] key
Filtering Range
• To change how much the scale fluctuates while reading, you can change
how often it filters, Enter the settings mode and it will display “F lL #”
• Then press [ZERO] to select the zero tracking range (1-6), 3 is the
default number. The higher the number the more stable the scale
• Finally, press [MODE] key to confirm it, go to the next menu
Zero Tracking
• To change how lenient the scale is at finding the zero point, In the set-
tings mode press [MODE] until the display reads “ZE #”
• Pres [ZERO] to select the speed (1-7) the default is 3, The bigger the
number the more lenient it is
• Finally, press [MODE] key to confirm it, go to the next menu
Linearity Correction
• To change the linearity calibration, Enter the settings mode and press
[MODE] until the screen reads “L Id #”
• Press [ZERO] to change the range (1-5), 5 being the default
• We suggest not to change the default settings
• Finally, press [MODE] key to confirm it, and the display will show the
next menu
Linearity Dot
• To change the correction point, Enter the settings mode and press
[MODE] until the display reads “L In #”
• Press [ZERO] to change the range (-8 - +8), 0 being the default
• We suggest not to change the default settings
• Finally, press [MODE] key to confirm it, go to the next menu
Selection of Maximum Capacity
• To change the max capacity, In the settings mode press [MODE] until
the display reads “C#####” ex. “C7.5000” means the capacity is 7.5kg
• The default number is the max capacity of the scale
• We suggest not to change the default settings
• Finally, press [MODE] key to confirm it, go to the next menu