Print Setup
There are three print types,
STABLE Print: print a stable reading when it is attained.
INSTANT Print: print immediately after press the Print key.
INTERVAL Print: print at predetermined time intervals.
Baud Rate Setup
Follow the procedures below:
1
.
Press Menu key.
2
.
Press Tare key, the balance reads “bAud”.
3
.
Press Count/Enter key, the balance reads “300”.
4
.
Press Tare key repeatedly, the balance display the other baud rates. Choose
the proper rate and press Count/Enter to confirm. The balance reads
“ParITy” and begins parity check.
5
.
Press Count/Enter key, the balance reads “nOnE” (no check) for the first.
6
.
Press Tare key, it reads the other check types, “Odd” indicates odd check
and “EvEn” indicates even check.
7
.
Choose the proper check type and press Count/Enter to confirm, the
balance now returns to weighing mode.
Enabling Units of Measure
The units function can be programmed to turn certain weighing units on or off. To
enable or disable certain units of measure, perform the following procedure.
1
.
Press Menu key, the display reads “PrInT”.
2
.
Press Tare key repeatedly until it reads “unIT”.
3
.
Press Count/Enter key, the display reads “g yes” which means g available for
use. Press Count/Enter to confirm. To disable g as the unit, press Tare key,
the display reads “g no”, press Count/Enter key to confirm.
4
.
Follow the steps above to enable or disable Oz, Ct or dwt as the unit of
measure.