5. Troubleshooting 

In the period of using microscope, if there is any trouble occurs, please referring to the following sheet 
listed some common troubleshooting resolve them.

 

Trouble 

Causation 

Remedy 

Bulb is dark when switch on 
power.     

Plug is unreliable 

Plug in again       

Bulb is broken 

Change bulb 

Fuse is broken 

Change fuse 

Bulb is flickering or 
brightness is unsteady. 
 

Bulb is unstable 

Insert it again 

Bulb is broken 

Replacing bulb 

Brightness of view field isn’t 
enough or is uneven. 

Bulb specification doesn’t   
meet the requirement. 

Replacing bulb 

Brightness isn’t adjusted correctly. 

Adjust potentiometer . 

Objective isn’t in correct position. 

Make  the  objective  in  correct 
position. 

The  size  of  aperture  diaphragm  is 
too small.   

Readjust aperture diaphragm . 

Brightness of view field isn’t 
enough or is uneven. 

Lens (objective, eyepiece, 
condenser, light collector) 

has dust. 

Clean it 

Position of condenser is too low. 

Higher condenser 

Image isn’t clear 
(contrast  or  definition  isn’t 
enough). 

Cover  glass  of  specimen  doesn’t 
meet the requirement. 

Use required thickness cover glass. 
(0.17mm) 

Cover glass of specimen isn’t in up 
direction. 

Place specimen correctly. 

Surface of objective lens 
Is dirty (especially it is easy for the 
front lens of 40X objective to dip in 
immersion oil). 

Clean it 

Immersion  oil  isn’t  used  for  100X 
objective (oil). 

Use immersion oil 

Immersion  oil  doesn’t  meet  the 
requirement. 

Use immersion oil supplied by us. 

There is bubble in immersion oil. 

Clear the bubble way. 

Size  of  aperture  diaphragm  isn’t 
right. 

Readjust aperture diaphragm. 

Position of condenser is too low. 

Readjust the position of condenser. 

One side of image is 
dark or image is 
moving as focusing.   

Objective isn’t in correct position. 

Make  the  objective  in  correct 
position. 

Specimen isn’t placed correctly 

Place specimen levelly on stage and 
clip it with clamp. 

Objective touches 
specimen  as  changing  low 
times objective to high times 
objective. 

Cover glass of specimen isn’t in up 
direction. 

Place specimen correctly. 

Cover  glass  doesn’t  meet  the 
requirement. 

Use required thickness 
cover glass (0.17mm). 

Image observed by two eyes 
aren’t 

in 

superposition 

entirely. 

Interpupilary distance isn’t adjusted 
correctly . 

Adjust interpupilary   
distance according to two eyes. 

It is easy for eyes to be tired 
during observing . 

Diopter isn’t adjusted correctly. 

Readjust diopter. 

 

Summary of Contents for A13.2604-A

Page 1: ...604 B Metallurgical Microscope Reflect Transmit Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly be fore start to use the instrument ...

Page 2: ...in use Alcohol gasoline paper and other flammable materials can t near the lamp in case of fire 5 Notes On Replacing The Bulb Replacement should be based on the identity of the instrument using the same specifications of the bulb otherwise it may cause equipment damage The power supply must be cut off before bulb replacement the bulb must be cooled off completely before proceeding 6 Carry Power mu...

Page 3: ...57X C mount also provide digital camera for CCD SLR digital camera or special designing camera by our company It has many advantages for example fine appearance stable performance excellent image convenient usage etc Widely used in factories universities and relative research institutes This series microscope is suitable for observing various metallurgical specimens casting quality and identificat...

Page 4: ...topf trinocular head with output connector for CCD Camera SLR Digital Camera 1 6 Nosepiece Quintuple 1 7 Mechanical stage For Reflecting Illumination Size 200mm 150mm Travel range 77mm 52mm For Transmitting Illumination Size175mm 145mm Travel range 77mm 52mm Two slid holder 1 8 Focusing systems Coaxial Coarse and Fine Focusing Knobs Coarse stroke 22mm Fine division 2µm Condenser up down range 25mm...

Page 5: ...Plate Stage Halogen lamp house Lamp house fix screw Reflection illumination Aperture diaphragm lever Filter slider Field diaphragm lever Polarizing slider Main body Right fine focusing knob Right coarse focusing knob Ground glass slider Slider handle Eyepiece tube fix screw Power line with plug Halogen Lamp Centering Knob MIT300 Upright Metallurgical Microscope ...

Page 6: ... switch Fuse Input socket Output socket MIT300 Upright Metallurgical Microscope Camera adaptor with dust cover for trinouclar head tube Handle Stage up down adjusting screw Stage plate fix screw Coarse tension adjusting knob Left coarse focusing knob Left fine focusing knob ...

Page 7: ...izer adjusting ring and fixed screw Field diaphragm adjusting ring Condenser fixed screw Potentiometer MIT500 Transmitting Reflecting Metallurgical Microscope Clamp locking screw Power switch Power socket and Fuse Adjustable Level for swing out Condenser Condenser focusing knob ...

Page 8: ...wer house if transmitting observing operate the potentiometer and power switch in the microscope main body 2 Sample placing Choose an available stage plate put the sample onto it turn the stage knob to put the sample into the light path If reflection observing stage plate and glass plate should be chosen if transmitting observing glass plate and clamp should be chosen 3 Put 10X objective into ligh...

Page 9: ...wn adjusting screw after get the available position 10 Usage of 100X immersion oil objective Adding moderate immersion oil between the front lens of 100X objective and the cover glass of specimen can make the image clearer Please pay attention that air bubble and impurity can t be in the immersion oil otherwise the image would be affected First of all take 40X objective which has been focused well...

Page 10: ...avoid fungus and mist damage instrument 4 2 4 Please pay attention to prevent microscope from violent shake and vibration in application and in carrying Don t drag it on the surface of worktable to avoid damage to microscope and worktable 4 3 Replacement of bulb 4 3 1 Turn off power and pull out plug 4 3 2 Wait the bulb become cool Please be sure that the bulb is cool then follow by the nest opera...

Page 11: ...lass 0 17mm Cover glass of specimen isn t in up direction Place specimen correctly Surface of objective lens Is dirty especially it is easy for the front lens of 40X objective to dip in immersion oil Clean it Immersion oil isn t used for 100X objective oil Use immersion oil Immersion oil doesn t meet the requirement Use immersion oil supplied by us There is bubble in immersion oil Clear the bubble...

Reviews: