29
4.3 Assembly for Episcopic Illumination Microscopy
4.3.1 Attaching the Filter Cube
1)
Remove the cover from the filter
cube mounting port.
2)
Turn the turret to a position where a
filter cube can be mounted easily.
3)
Check the address on the turret, and
align the filter cube with the guide
groove before inserting.
4)
Cover with cover.
4.3.2 Installing the Fluorescence LED
Unit
1)
Loosen the four screws fixing the
cover of the fluorescence LED unit
on the back of the microscope, and
remove the cover.
2)
Mount the LED unit (corresponding
to the desired wavelength) at the
position that matches the address of
the filter cube. The back cover of the
LED unit is marked with position
number. The corresponding position
is as below:
Sign
LED light
1
G light
2
B light
3
U light
3)
Remount the cover in the original
position.
4.3.3 Replacing the Fuse
Open the cover, and in the fluorescence
LED unit can see two pieces of fuse (as
shown in the right figure), with a slotted
screwdriver to counterclockwise unscrew
the fuse.
★
The middle of the fuse is thin glass,
please be careful when changing.
1
2
Filter cube
Cover
1
3
2
Fuse
LED light source
LED unit cover
1
2