2

Eyepiece

1

 

Type 

Magnification 

Diameter of Viewing Field(mm) 

Quantity 

Wide Field Eyepiece 

10 

18 

Wide Field Eyepiece 

16 

13 

(Optional) 

3

Viewing Head

2

 

    It is a monocular, or dual, or binocular or trinocular head. Binocular or trinocular head is 
separated from the microscope body for transport safety. It is only fixed on the analyzer 
part  with  screw  (28)  before  using.  It  is  30°  /45°  inclined  and  360°  rotatable.  Its 
interpupillary  distance  can  be  adjusted  from  55mm  to  75mm  to  suit  for  the  different 
viewers. There is a rod for prism transition in the trinocular head. It is used to change the 
distribution of light. a) 

In: 

Light is totally transmitted in binocular tube, to ensure enough 

brightness when viewing. b) 

Out:

 The Spectroscope built-in divides the light into two parts. 

This  moment,  you  may  use  the  device  in  the  vertical  tube,  such  as  CCD  camera, 
electronic eyepiece, etc.

 

4

Analyzer Part

 

    It is comprised of analyzer and compensator. It is separated from the microscope body 
for transport  safety.  It  is only  fixed  on  the microscope  body  (16) with  screw (25)  before 
using.

   

When  the  microscope  is  only  used  for  biological  microscopy  observation,  put  the 

compensator (26) on the middle position, and move the analyzer (27) out . 

Insert the analyzer(27) in and drive the analyzer around the optical axis. Scale on the 

graduated ring and the vernier will tell you the angle analyzer turned. 

You may select λ Slip (first class red)

or 1/4λ Slip to finish your special task.

   

5

Objective

4

 

Normally

the objectives in this microscope are non-stress achromatic objectives: 

Magnification  Numerical Aperture(N.A.) 

Thickness of Cover Glass(mm) 

Quantity 

0.10 

 

10 

0.25 

 

40 

0.65 

0.17 

100 

1.25 

0.17 

Total magnification after grouping eyepieces and objectives

 

                                      Objective 

Total Magnification

 

Eyepiece 

10 

40 

100 

WF10X 

40 

100 

400 

1000 

WF16X 

64 

160 

640 

1600 

6

Round Stage

6

  

  It is a revolving round stage, its diameter is 120mm, and it can 360

0

 rotate freely. It is 

graduated in 360

area with

 

1

0

 minimum increments. When vernier scale (7) is used, the 

revolving angle precision of the stage is 0.1

0

.   

   

P.3 

Summary of Contents for A15.1302

Page 1: ...A15 1302 Polarizing Microscope Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly be fore start to use the instrument...

Page 2: ...doing so in the manual please don t dismantle your microscope Otherwise it will damage the microscope seriously and reduce greatly its accuracy and using life When you identify some troubles and can...

Page 3: ...P 1 B Structure and Specification of Microscope P 1 1 Picture of Microscope P 1 2 Structure Diagram of Microscope P 2 3 Functions and Specifications of Parts P 2 1 Micro camera Components Option Acce...

Page 4: ...unction It can observe mineral crystal figure color and interference color and also identify its optical performance It is a good microscope used in geology petroleum coal Chemical fiber medical treat...

Page 5: ...Stage 16 Body 17 Plastic Cover 18 Rubber Foot 19 Brightness Adjusting Plate 20 Focusing Tension Knob 21 Coarse Focusing Knob 22 Fine Focusing Knob 23 Knob for Condenser Up Down 24 ABBE Condenser 25 Lo...

Page 6: ...t is separated from the microscope body for transport safety It is only fixed on the microscope body 16 with screw 25 before using When the microscope is only used for biological microscopy observatio...

Page 7: ...te 13 the deflexion angle of the polarizer will change Scale on the graduated ring will tell you the angle the polarizer turned 10 Illumination System The input voltage for the microscope may be 220V...

Page 8: ...X 13 Wide field Eyepiece Optional 7 Slip first class red 1Pcs 8 1 4 Slip 1Pcs 9 Blue filter 1Pcs 10 Fuse 1Pcs 11 Dust cover 1Pcs 12 Service Manual 1Copy 13 Spare bulb 1Pcs Only for Halogen Lamp D How...

Page 9: ...viewing field 7 How to observe in orthogonal polarizing condition 1 Make the analyzer is orthogonal with the polarizer The analyzer is in south north and the polarizer is in east west 2 The condenser...

Reviews: