2
)
Eyepiece
(
1
)
:
Type
Magnification
Diameter of Viewing Field(mm)
Quantity
Wide Field Eyepiece
10
18
2
Wide Field Eyepiece
16
13
(Optional)
3
)
Viewing Head
(
2
)
:
It is a monocular, or dual, or binocular or trinocular head. Binocular or trinocular head is
separated from the microscope body for transport safety. It is only fixed on the analyzer
part with screw (28) before using. It is 30° /45° inclined and 360° rotatable. Its
interpupillary distance can be adjusted from 55mm to 75mm to suit for the different
viewers. There is a rod for prism transition in the trinocular head. It is used to change the
distribution of light. a)
In:
Light is totally transmitted in binocular tube, to ensure enough
brightness when viewing. b)
Out:
The Spectroscope built-in divides the light into two parts.
This moment, you may use the device in the vertical tube, such as CCD camera,
electronic eyepiece, etc.
4
)
Analyzer Part
:
It is comprised of analyzer and compensator. It is separated from the microscope body
for transport safety. It is only fixed on the microscope body (16) with screw (25) before
using.
When the microscope is only used for biological microscopy observation, put the
compensator (26) on the middle position, and move the analyzer (27) out .
Insert the analyzer(27) in and drive the analyzer around the optical axis. Scale on the
graduated ring and the vernier will tell you the angle analyzer turned.
You may select λ Slip (first class red)
,
or 1/4λ Slip to finish your special task.
5
)
Objective
(
4
)
:
Normally
,
the objectives in this microscope are non-stress achromatic objectives:
Magnification Numerical Aperture(N.A.)
Thickness of Cover Glass(mm)
Quantity
4
0.10
1
10
0.25
1
40
0.65
0.17
1
100
1.25
0.17
1
Total magnification after grouping eyepieces and objectives
:
Objective
Total Magnification
Eyepiece
4
10
40
100
WF10X
40
100
400
1000
WF16X
64
160
640
1600
6
)
Round Stage
(
6
)
:
It is a revolving round stage, its diameter is 120mm, and it can 360
0
rotate freely. It is
graduated in 360
0
area with
1
0
minimum increments. When vernier scale (7) is used, the
revolving angle precision of the stage is 0.1
0
,
.
P.3