10
-
6. Use of fluorescent centering device
The Fluorescent centering device can center
directly the light source for convenient using.
The size of the device is similar to the normal
objective, and there is an observation window in
the side of cover.
a.
Hold the silvery colour gear in the centering
device, and install it on the nosepiece ,turn
the cover of centering device to align the
observation window .
b.
Turn the fluorescence device to G, and watch
the light source’s position through the widow.
c.
Adjust the handle of light source condenser properly, and make the outline of
the light source clear, adjusting the knob on fluorescence lamp house to
center.
d.
The cross line center in observation window must be aligned with the lamp
house .
▲
Please turn the fluorescence device to G, because the brightness of G is
suitable for observation better to avoid to feel unconformable for high
brightness.
▲
Maybe can't get better image afer adjustment ,you need to fine adjusting
again in normal specimen observation.
observation
window