11
4. specification A24.1001
Specification
Zoom ratio
1
:
6.3
Zoom magnification
0.8
×~
5
×
Binocular
viewing
head
Inclined at 45
°
interpupillary distance range: 52
~
75mm
Working distance
115mm
Eyepiece
15
×
/
Φ
16 with diopter adjustment
Focusing range
110mm
Illumination system
6V/20W Halogen illumination
8W fluorescent lamp
LED lighting
Stand
0-38
°
Rotatable
Base
325
°
Rotatable
5.
Troubleshooting
Under certain conditions, performance of this unit may be adversely affected by factors other
than defects. If a problem occurs, please review the following list and take remedial action as needed.
If you cannot solve the problem after checking the entire list, please contact your local franchiser for
assistance.
PROBLEM
CAUSE
SOLUTION
PAGE
1
.
Incomplete binocular
vision.
Interpupillary distance is
not correctly adjusted.
Adjust it correctly
8
Diopter adjustment is
incomplete.
Complete
diopter
adjustment
2
.
Dirt is visible in field of
view
Dirt on specimen
Clean specimen
2
Dirt on eyepiece
Clean eyepiece
3
.
The visibility of the
image is poor.
Dust on objective front
lens.
Clean lens surface