background image

 

12 

PROBLEM 

CAUSE 

SOLUTION 

PAGE 

4. Specimen image blurs 

when zoom magnification is 

changed. 

Diopter ring is not 

correctly adjusted. 

Adjust it correctly. 

Not in complete focus on 

specimen. 

Focus specimen 

correctly at a high 

magnification. 

5

Coarse focus adjustment 

knobs rotate with too much 

resistance. 

Tension adjustment ring 

is too tight. 

Loosen it properly. 

6

Zoom microscope body 

drops or specimen goes out 

of focus during observation.   

Tension adjustment ring 

is too loose. 

Tighten it properly. 

 

6

Outfit

 

 

Components 

Specification 

Quantity 

NGI-6 

Main body 

Main body 

 

Observation 

system 

Binocular head and zoom body 

 

Trinocular head 

 

Illuminator 

Transmitted illuminator (Halogen illumination) 

 

Reflected 

illuminator 

fluorescent lamp 

 

LED lighting 

 

Eyepiece 

15

×

Eyepiece 

 

20

×

Eyepiece 

 

Objective 

0.5

×

 

 

0.7

×

 

 

2

×

 

 

Stone holder 

Stone holder 

 

Photo attachment 

 

 

Video attachment 

 

 

Power cord 

Power Cord 

 

●:

standard Outfit

;○:

Optional 

Summary of Contents for A24.1001

Page 1: ...croscope Instruction Manual A24 1001 It is recommended strongly that you study this manual thoroughly before using the microscope Retain this manual in an easily accessible place near the work desk for future ...

Page 2: ... Assembly 4 2 1 Assembly Diagram 4 2 2 Detailed Assembly Procedure 5 3 Adjustment and Operation 7 3 1 Stand Adjustment 7 3 2 Adjusting Binocular Viewing Head 8 3 3 Bottom Illumination 9 3 4 Top Illumination 10 4 Specification 11 5 Troubleshooting 11 6 Outfit 12 ...

Page 3: ...c shock 7 Use the factory supplied power cord please 2 Maintenance and Storage 1 All the lenses have been well checked and adjusted It is forbidden to disassemble them yourself 2 The binocular viewing head and focusing unit have a compact and precise frame please don t disassemble them as possibly as you can 3 Keep the instrument clean wipe dust regularly and be attention to avoid contaminating th...

Page 4: ...3 Components A24 1001 1 Name of Components A24 1001 15 Eyepieces Stand Focusing knobs Base Main Switch Fluorescent lamp LED lighting Stone Holder Zoom body knob Zoom body Brightness adjusting knob ...

Page 5: ... shows how to assemble the various modules The numbers indicate the order of assembly When assembling the microscope make sure that all parts are free of dust and dirt and avoid scratching any parts or touching glass surfaces A24 1001 Binocular head locking bolt ...

Page 6: ...hows Keep your eyes at the correct distance from eyepieces and make the microscope even more comfortable to use 2 2 3Install stoneholder Fig 2 The stage has two stoneholder mounts where you attach the stoneholder Note Working Environment Requirement 1 Ambient temperature 5 C to 40 C 41 F to 104 F Maximum R H 85 2 High Temperature will result in a mildewing dew and even ruinous instrument 3 Avoid p...

Page 7: ...age and seat it as deeply as possible 2 Tightening the screw cap to fix it 3 The tile of fluorescent lamp illumination is adjustable for individual comfortable 2 2 5 Install LED lighting Fig 4 1 Insert LED pipe into the socket and seat it as deeply as possible 2 Remove the plastic cover of the LED pipe Fig 4 Fig 3 ...

Page 8: ...ke hold of bracket with right hand then push or pull the bracket to obtain the most comfortable viewing position The stand is 0 38 adjustable 3 1 3 Adjusting the rotation tension of the focus adjustment knob Fig 6 1 Rotating the focus adjustment knob to lower the zoom body to a lower place for fear that the zoom body drops down suddenly if the tension is decreased too much 2 Hold the focus adjustm...

Page 9: ... the iris diaphragm by turning the iris diaphragm control lever clockwise 2 Set the dioptor rings of both eyepieces to 0 position Do this when users change because different users have different diopters 3 Rotate the zoom body knob to the highest magnification 5 and turn the focusing knob to focus the iris diaphragm 4 Rotate the zoom body knob to the lowest magnification 0 8 adjust the diopter rin...

Page 10: ...ragm control lever 2 Brightfield illumination open the baffle by turning the baffle control knob counter clockwise Inclusions should stand out as dark objects against a bright background This method helps you see curved striae or other low relief or structural inclusions 3 Darkfield illumination close the baffle by turning the baffle control knob clockwise the stone is illuminated with diffused li...

Page 11: ...adjustable for individual comfort 2 LED lighting turning the main switch to I on station after installed the LED lighting and rotate the left side brightness adjusting knob to obtain a comfortable brightness It can offer many lighting options different types of lighting work best for seeing different gem features Brightfield darkfield top illumination these three types of illumination is optional ...

Page 12: ...ay be adversely affected by factors other than defects If a problem occurs please review the following list and take remedial action as needed If you cannot solve the problem after checking the entire list please contact your local franchiser for assistance PROBLEM CAUSE SOLUTION PAGE 1 Incomplete binocular vision Interpupillary distance is not correctly adjusted Adjust it correctly 8 Diopter adju...

Page 13: ... microscope body drops or specimen goes out of focus during observation Tension adjustment ring is too loose Tighten it properly 7 6 Outfit Components Specification Quantity NGI 6 Main body Main body 1 Observation system Binocular head and zoom body 1 Trinocular head 1 Illuminator Transmitted illuminator Halogen illumination 1 Reflected illuminator fluorescent lamp 1 LED lighting 1 Eyepiece 15 Eye...

Reviews: