5.2.4.4 Adjust the electrical alignment
The electron beam can be modulated in the electrical alignment. When the
filament works for a while, the electron beam may have some deviation. Some
corrections can be made by adjusting Beam tilt and Beam shift in the software.
(this feature does not need regular adjustment) Make ensure that the alignment of
final aperture is normal; Click the filament image function; Adjust Beam tilt and
Beam shift with the slider, and you will see the circular filament spot moving with
the adjustment. Move it to the center of the screen. It is necessary to adjust the
final aperture alignment and eliminate astigmatism..
5.2.5Sample preparation
The sample preparation of scanning electron microscopy is very simple. A piece of
sample can be cut off and pasted onto the sample cup with conductive adhesive glue
or tape. If the sample is not conductive, stick the sample on the sample cup with
conductive glue or conductive tape, and then sputter a layer of conductive film
(usually gold or carbon) with sputtering apparatus. Please refer to the sputtering
apparatus operation manual for specific operation method
5.2.6 Image processing operation
Open the image processing window, you can do a variety of image processing
operations..
Summary of Contents for A63.7080
Page 1: ...A63 7080 81 Scanning Electron Microscope SEM Instruction Manual ...
Page 17: ......
Page 20: ...Fig 3 7 The function button and state display of the scan control window ...
Page 26: ...4 Automatic sample stage 4 1Sample stage control interface ...
Page 35: ...Function The following table contains information on the components of the Settings tab ...