optris CTratio
– E2013-04-A
12
Sensor placement [1C mode]
The size of the measuring object and the optical resolution of the infrared thermometer determine the
maximum distance between sensing head and measuring object. In order to prevent measuring errors the
object should fill out the field of view of the optics completely. Consequently, the spot should at all times have
at least
the same size
like the object or should be
smaller than
that.
Sensor placement [2C mode]
In the 2-color-mode the sensing head can be placed under various conditions, like:
Measurement through holes smaller than the spot size
Measurement through dust, smoke or steam in the atmosphere
Measurement of objects smaller than the measurement spot
Measurement through a dirty lens or dirty sighting window
[► The Ratio Principle]
Please note that the sensor accuracy is not guaranteed for signal reductions of more than 95%.