optris CTratio
– E2013-04-A
17
Laser Sighting
The integrated laser sighting supports the alignment and focusing of the optics. The size
of the laser dot is equal to the real measurement spot size at any distance.
If the laser is activated the temperature measurement is not possible!
The laser can be activated/ deactivated via the programming keys on the unit or via the
software. If the laser is activated a yellow LED will shine (beside temperature display).
[
► Electrical Installation/ Operating]
At ambient temperatures >70 °C on the electronic box the laser will switch off
automatically.
WARNING: Inside the electronic box a HIGH POWER LASER is located. The laser light is coupled into
the fiber optics. On the outlet at the optical head the laser power is < 1mW and fulfils laser class 2
according IEC 60825-1:2007.
Before you open the box, remove the fiber optical cable from the electronics or disassemble the
optical head the sensor must be switched off and disconnected from the power supply and USB
interface!
SAFETY SWITCH: If the fiber optical cable will be removed from the electronics the laser will be
switched off automatically. Any manipulation at this safety switch is prohibited.
WARNING: Do not point the laser directly at the eyes of persons or animals! Do not stare into the laser
beam. Avoid indirect exposure via reflective surfaces!