16
8.1.3 Calibration of Resistances
Connect Ohm-Meter to the calibrator outputs. The required accuracy is 0.05%.
8.2
CALIBRATION MULTIMETER
Select the required range (2) or (3).
The Range (2) is for calibration of the current, the Range (3) is for the voltage calibration.
The key
ESC
terminates the calibration and the display returns into the measuring mode.
Set with the keyboard the resistance value which is being
measured with the connected Ohm-Meter.
Set with the keyboard the resistance value which is being
measured with the connected Ohm-Meter.
Set with the keyboard the resistance value which is being
measured with the connected Ohm-Meter.
Set with the keyboard the resistance value which is being
measured with the connected Ohm-Meter.
Supply from external current calibrator 0.000 mA and confirm with
OK. The point 0.00A is calibrated.
Supply from external current calibrator 100.000 mA and confirm
with OK. The point 0.1A is calibrated.
Supply from external voltage calibrator 0.000 V and confirm with
OK. The point 0V is calibrated.
Supply from external voltage calibrator 2.000 V and confirm with
OK. The point 2V is calibrated
Supply from external voltage calibrator 20.000 V and confirm with
OK. The point 20V is calibrated.
Supply from external voltage calibrator 0.000 V and confirm with
OK. The point 0V is calibrated.
Supply from external voltage calibrator 0.000 V and confirm with
OK. The point 0V is calibrated.
.
Supply from external voltage calibrator 200.000 V and confirm with
OK. The point 200V is calibrated.
Summary of Contents for OC 505
Page 2: ...2 ...