56
|
INSTRUCTIONS FOR USE
OM 402LC
6.4.3
Calibration - Input range
6.4.4
Selection of instrument menu language version
CHANNE.
SERVIC.
OUTPUT.
MENU
RESTOR.
N. PASS.
INPUTS
IDENT.
CALIB.
C. MIN
YES
C. MAX
LANG.
CALIB.
Input range
calibration
- prior performing any changes you will be
asked to confirm your selection “YES”
C. MIN
Calibration of the
beginning of the
measuring range
- prior confirmation of the selection the
reference signal has to be connected
C. MAX
Calibration of the end of
the measuring range
- prior confirmation of the selection the
reference signal has to be connected
CHANNE.
SERVIC.
OUTPUT.
MENU
RESTOR.
N. PASS.
INPUTS
IDENT.
CALIB.
LANG.
ENGL.
CZECH
DEF
LANG.
Selection of instrument
menu language version
CZECH
Instrument menu is in
Czech
ENGL.
Instrument menu is in
English
profi
profi
6
SETTING
After incorrect client calibration it is always possible to restore
manufacture calibration (“SERVIC./RESTOR/CALIB.”)
!
Manual calibration:
MAX Sensor
range
SENSE Sensor
sensitiveness
Automatic calibration
(after calibartion in menu “SERVIC./CALIB.”):
MIN
Size of load with which minimum
calibration was performed
MAX
Size of load with which maximum
calibration was performed
- upon maximum calibration we recom-
mend the reference load value in the upper third of
the measuring range
!