8. ERROR STATEMENTS
ERROR
CAUSE
ELIMINATION
E. D.U.
display underflow
number is too small (large negative) to be
displayed
change display projection
E. D.O..
display overflow
number is too large to be displayed
change display projection
E. T.U.
table underflow
number outside tablerange
change display projection or table range
E. T.O.
table overflow
number outside tablerange
change display projection or table range
E. I.U.
range underflow (A/D converter)
input signal is smaller than the one allowed
by instruments range
change input signal value
E.I.O.
range overflow (A/D converter)
input signal is larger than the one allowed
by instruments range
change input signal value
E.HW.
some part of the instruments does not
function properly
when reported repeatedly send the instru-
ment for repair
E.EE.
EEPROM error
damaged data in EEPROM
when reported repeatedly send the instru-
ment for repair
E.DT.
DATA error
data in EEPROM outside the range
information about performed automatic
data correction in EEPROM
E.CL.
EEPROM error
memory is blank („DEF“ values presetting
has initialized)
„Def“ values will be used in emergency
but calibration data will be impaired
> send for repair