DP-100SL
23
Site Preparation Guide
Chapter 4
Environment & Climate
Clean Room
Clean room level class 100K or better (class 10K is recommended)
according to US-Federal-Standard 209E is required in order to reach the
specified system specifications. No organic chemicals or ammonia may be
present in the air of the clean room.
We strongly recommend to place the DP-100SL in an independent room
environment (i.e. not be part of a larger room), with the following
conditions:
• Ceiling-to-floor (laminar) filtered airflow to reduce particulate
contamination.
• Class 10K clean room (per Fed Spec 209E).
• Inline active carbon filtration system.
• Dedicated standalone air makeup system that is free from organic
chemicals and ammonia that may attack system components.
Climate Control
Room climate control must be designed so that operating temperatures
and humidity, as described in
, are
maintained when the system operates at full power (see
). Do not supply electrical power for the climate control system
from the same source as the system.