24
Chapter 4
Environment & Climate
The customer needs to install temp/humidity meter in the room.
Room Temperature
*
Storage temperature allowed only when in original packaging.
Vacuum room
Vacuum room must be ventilated. Temperature in the vacuum room
should not exceed 35 ºC.
!
Caution
The air flow should not be directed toward the machine.
Table 4.1 Operating and Storage Temperatures
Unit
Operating Temperature
Storage Temperature*
DP-100SL
20
°
C to 23
°
C (68
°
F to 73
°
F)
0
°
C to 35
°
C (32
°
F to 95
°
F)
!
Caution
In order to achieve the specified system performance, the imager room
temperature must be set within the above range, and must not fluctuate
more than ±2 ºC from the set point during operational hours.