DP-100SL
25
Site Preparation Guide
Humidity
* Storage humidity allowed only when in original packaging.
Heat Dissipation
Ambient Air Conditions
You should install:
• A fresh air outlet above the laser supply box (i.e. power heat
sink).
• A warm air exhaust close to the main heat source (i.e. laser supply box).
Take note that:
• No organic chemicals allowed in the air.
• No ammonia allowed in the air.
Table 4.2 Operating Humidity
Unit
Operating Humidity
Storage Humidity*
DP-100SL
50% to 60%
10% to 80%, non-condensing
Table 4.3 Maximum Heat Dissipation for System Units
Unit
Heat Dissipation
DP-100SL
5 kW (228 BTU/minute) - To air
Pump
2.5 kW (114 BTU/minute) - To air
Summary of Contents for DP-100SL
Page 1: ...DP 100SL Site Preparation Guide October 2003 Revision A ...
Page 2: ......
Page 14: ......
Page 18: ......
Page 25: ...DP 100SL 19 Site Preparation Guide Figure 3 2 System Dimensions Front View ...
Page 26: ...20 Chapter 3 Facility Physical Requirements Figure 3 3 System Dimensions Side View ...
Page 27: ...DP 100SL 21 Site Preparation Guide Figure 3 4 Site Dimensions Standard Configuration ...
Page 32: ......
Page 40: ......
Page 48: ......