Site Preparation Guide
Chapter 5
Electricity
Power Requirements
* Recommended current specifications for electrical outlet.
Table 5.1 Power Requirements for the DP-100SL machine
DP-100SL, 3
Φ
+ n ground
400 V
(Europe)
480 V
(USA)
200 V
(Japan)
Voltage
400 V+6% /-10%
480 V
±
10%
200 V
±
10%
Frequency 50
±
1 Hz
60
±
1 Hz
50/60
±
1 Hz
Current*
3 x 16 A
3 x 16 A
3 x 32 A
Maximum power consumption
6 kVA
6 kVA
6 kVA
Note
• The DP-100SL machine comes with a 10m connecting cable, that must be
connected to a general clamp at the rear wall close to the DP-100SL.
• The transformer , when required, is located inside the DP-100SL machine.
• An additional ground connection must be installed in parallel to the machine’s
power connecting cable (the machine comes with an additional 10m cable, as
mentioned). The earth-connection must be done at same point as for the
earthing wire of power.