28
Chapter 5
Electricity
* Recommended current specifications for electrical outlet.
Grounding
The DP-100SL ground should be connected to a special ground line, which
is connected to the building ground.
A fixed installation to the ground
has to be provided before switching on the system.
Ground
connection impedance must be <0.5 Ohms.
Voltage Transients
show allowable voltage transients. Those should be
tested when all machinery and air conditioning units are operating
normally and the system is turned on.
Table 5.2 Power Requirements for the Pump
Pump, 3
Φ
+ ground
400 V
(Europe)
208 V
(USA)
200 V
(Japan)
Voltage
400 V+6% /-10%
208 V
±
10%
200 V
±
10%
Frequency 50
±
1 Hz
60 Hz
50/60 Hz
Current*
3 x 16 A
3 x 16 A
3 x 16 A
Maximum power consumption
2.5 kVA
2.5 kVA
2.5 kVA
Note
An additional ground connection must be installed in parallel to the machine’s
power connecting cable (the machine comes with an additional 10m cable, as
mentioned). The earth-connection must be done at same point as for the
earthing wire of power.
Summary of Contents for DP-100SL
Page 1: ...DP 100SL Site Preparation Guide October 2003 Revision A ...
Page 2: ......
Page 14: ......
Page 18: ......
Page 25: ...DP 100SL 19 Site Preparation Guide Figure 3 2 System Dimensions Front View ...
Page 26: ...20 Chapter 3 Facility Physical Requirements Figure 3 3 System Dimensions Side View ...
Page 27: ...DP 100SL 21 Site Preparation Guide Figure 3 4 Site Dimensions Standard Configuration ...
Page 32: ......
Page 40: ......
Page 48: ......