DP-100SL
29
Site Preparation Guide
Unless the following specs can be met by current on-site electrical
installations, a line conditioner is required.
Table 5.3 Allowable Voltage Transients–USA
Amplitude
Measured At
±
10%
Hot to Neutral
0.5 V
Neutral to Ground (common mode)
Table 5.4 Allowable Voltage Transients–Europe/Japan
Transient Duration
(msec)
Transient Amplitude (Volts)
Japan
Europe
<5
<55
<110
5 to 50
<25
<50
>50
±
10% nominal
(according to
)
±
10% nominal
(according to
)
Summary of Contents for DP-100SL
Page 1: ...DP 100SL Site Preparation Guide October 2003 Revision A ...
Page 2: ......
Page 14: ......
Page 18: ......
Page 25: ...DP 100SL 19 Site Preparation Guide Figure 3 2 System Dimensions Front View ...
Page 26: ...20 Chapter 3 Facility Physical Requirements Figure 3 3 System Dimensions Side View ...
Page 27: ...DP 100SL 21 Site Preparation Guide Figure 3 4 Site Dimensions Standard Configuration ...
Page 32: ......
Page 40: ......
Page 48: ......