44
Appendix B
Customer Site Preparation Checklist
Electrical Issues
Complete
Pending
Date
Notes
Requirements for the DP-100SL machine
(as listed in
Main electrical power supply has a switch.
Maximum voltage transient of _____________
Automatic slow blow circuit breaker installed in
series in each power line.
DP-100SL connected to special ground line which is
connected to the building ground.
Ground connection impedance less than 0.5 Ohms.
Electrical power for air conditioning is from
different source than the system.
Requirements for the Pump
(as listed in
Imaging Room Issues
Complete
Pending
Date
Notes
Image room size at least:
3.57W x 5.30L x 2.80H (meters)
11.7W x 17.4L x 9.2H (feet)
The door is at least 1800 mm (71”) wide
and 2010 mm (79”) high.
Room isolated to safeguard against laser hazards.
Room doors have a lock that can be locked from
the inside.
A box with a key for emergency entry is located
outside the room.
Paint in room is non-shedding and washable.