background image

6

Do not loosen the center screw. The goal is to center the pri-
mary mirror reflection in the secondary mirror, as in Figure 
5d. Don’t worry that the reflection of the secondary mirror (the 
smallest circle, with the collimation cap “dot” in the center) is 
off-center, since that adjustment is made when aligning the 
primary mirror in the next step

Aligning the Primary Mirror

The final adjustment is made to the primary mirror. It will need 
adjustment if, as in Figure 5d, the secondary mirror is cen-
tered under the focuser and the reflection of the primary mir-
ror is centered in the secondary mirror, but the small reflection 
of the secondary mirror (with the “dot” of the collimation cap) 
is off-center.
The tilt of the primary mirror is adjusted with three pairs of 
collimation screws (Figure 9). The collimation screws can be 
turned with a Phillips head screwdriver and a 2.5mm hex key. 
Each pair of collimation screws work together to adjust the tilt 
of the primary mirror. The allen-head screw pushes the mir-
ror forward while the Phillips head screw pulls the mirror cell 
back. One must be loosened and the other tightened by the 
same amount in order to adjust the tilt.
To adjust the primary mirror tilt:
1.  Tighten and loosen one of the pairs of collimation screws 

by one turn.

2.  Look into the focuser using the collimation cap and see if 

the secondary mirror reflection has moved closer to the 
center of the primary. You can tell this easily with the col-
limation cap and mirror center mark by simply watching 
to see of the “dot” of the collimation cap is moving closer 
or farther away from the ring on the center of the primary 
mirror.

3.  Repeat this process on the other two pairs of collimation 

screws, if necessary. It will take a little trial and error to get 
a feel for how to adjust the mirror to center the “dot” of the 
collimation cap in the ring of the mirror mark.

When you have the dot centered as much as possible in the 
ring, your primary mirror is collimated. The view through the 
collimation  cap  should  resemble  Figure  5d.  Make  sure  all 
the collimation screws are tight (but do not over tighten), to 

secure the mirror tilt. A simple star test will indicate how well 
the telescope optics are collimated.

Star-Testing the Telescope

When it is dark, point the telescope at a bright star and accu-
rately center it in the eyepiece’s field of view. Slowly de-focus 
the image with the focusing knob. If the telescope is correct-
ly collimated, the expanding disk should be a perfect circle 
(Figure 10). If the image is unsymmetrical, the scope is out 
of collimation. The dark shadow cast by the secondary mirror 
should appear in the very center of the out-of-focus circle, like 
the hole in a donut. If the “hole” appears off-center, the tele-
scope is out of collimation.
If you try the star test and the bright star you have selected is 
not accurately centered in the eyepiece, the optics will always 
appear out of collimation, even though they may be perfectly 
aligned. It is critical to keep the star centered, so over time you 
will need to make slight corrections to the telescope’s posi-
tion in order to account for the sky’s apparent motion. Point 
the telescope at Polaris (the north star) if you do not have a 
mount that tracks.

Lateral Support System

One of the exciting new features of this telescope is lateral 
support screws to secure the position of both the primary mir-
ror and corrector plate which will further stabilize the optics. 
These screws are visible around the perimeter of the tube on 
both front and back (Figures 11a and 11b) and have been 
set at the factory. The Phillips-head screws pictured in Figure 
11b are cover screws. The Lateral Support screws can be 
accessed by adjusting these screws (Figure 11c). No adjust-
ment of this component is necessary when performing routine 
collimation. If larger adjustments do need to be made simply 
loosen these support screws slightly. When finished re-tighten 
just tight enough that you feel the slightest resistance when 
tightening.  Do  not  over-tighten  these  screws  as  pinched-
optics or other distortions could result. 

Care & Maintenance

Give your telescope reasonable care and it will last a lifetime. 
When not in use, keep its dust cover on as well as the dust 

Figure 10. 

A star test will determine if a telescope’s optics are 

properly collimated. An unfocused view of a bright star through 

the eyepiece should appear as illustrated on right if optics are 

perfectly collimated. If circle is unsymmetrical, as in illustration on 

left, scope needs collimation.

Out of collimation

Collimated

Figure 9. 

The 

tilt of the primary 

mirror is adjusted 

with three pairs 

of collimation 

screws.

Primary collimation 
adjustment screws

Summary of Contents for 9978

Page 1: ...Support 800 676 1343 E mail support telescope com Corporate Offices 831 763 7000 89 Hangar Way Watsonville CA 95076 Orion Premium 190mm f 5 3 Maksutov Newtonian Astrograph 9978 Providing Exceptional C...

Page 2: ...2 accessories including just about any eyepiece or camera The Crayford focuser has a 2 eyepiece holder with a 1 25 eyepiece adapter installed Both the 2 eyepiece holder and 1 25 eyepiece adapter feat...

Page 3: ...elescope is optimized for use with an APS size sensor such as the Orion StarShoot Pro or a DSLR camera Cooling the Telescope All optical instruments need time to reach thermal equil ibrium The bigger...

Page 4: ...tension to the focuser drawtube Figure 2 will increase the amount of weight the focuser can handle Install your camera onto the supplied camera adapter on the focuser during the day Check the focuser...

Page 5: ...reflections for the time being The secondary mirror itself should be cen tered in the focuser drawtube in the direction parallel to the length of the telescope The secondary mirror should face the foc...

Page 6: ...ope When it is dark point the telescope at a bright star and accu rately center it in the eyepiece s field of view Slowly de focus the image with the focusing knob If the telescope is correct ly colli...

Page 7: ...compressed air can or blower bulb available from Orion Any remaining dust is best left alone unless the build up is extreme Finger prints and water marks should be cleaned from your telescope s corre...

Page 8: ...turned postage paid to Orion Warranty Repair 89 Hangar Way Watsonville CA 95076 Proof of purchase such as a copy of the original receipt is required This warranty does not apply if in Orion s judgment...

Reviews: